Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Empower User Question

Empower Tip: Empower User Question

Tip #234: Empower User Question: Reporting Peaks Below Limit of Quantitation (LOQ) and/or Limit of Detection (LOD) using a Custom Field in Empower

Tip #234: Empower User Question: Reporting Peaks Below Limit of Quantitation (LOQ) and/or Limit of Detection (LOD) using a Custom Field in Empower

In this tip, we will answer an Empower user question.

QUESTION:

Can Empower report if peaks are below the Limit of Quantitation (LOQ) and/or below the Limit of Detection (LOD) using a Custom Field?

The answer is YES. I will explain how Empower can report if peaks are below the Limit of Quantitation and/or below the Limit of Detection using a Custom Field. Classically, Limit of Quantitation is defined at a signal-to-noise ratio of 10 and Limit of Detection is defined at a signal-to-noise ratio of 3.

STEP 1

STEP 1

There are ‘Special Values’ which are used in Custom Fields. Using these ‘Special Values’ will either return a calculated value, such as ‘Amount’, or if the calculation results in -60002 then Empower will return the answer ’Not Detected’. In this way, you can get both calculated results together with text results..

STEP 2

STEP 2

Our example could result in 1 of 3 possible outcomes. The portion of the equation in ‘green’ will result in Empower reporting the ‘Calculated Amount’, the portion in ‘red’ will result in Empower reporting ‘Below Detection Limit’ and the portion in ‘orange’ will result in Empower reporting ‘Below Peak Quantitation Limit’.

STEP 3

STEP 3

The peak Custom Field uses Data Type, ‘Real’ and the Boolean expressions are joined using the ‘+’.

STEP 4

STEP 4

The result displays one peak ‘Below Peak Quantitation Limit’ while the other peaks report the ‘Calculated Amount’

It's that easy!

Final Notes

You can use either the Pro or QuickStart interface for this tip.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



Benötigen Sie Hilfe? Kontakt mit einem Waters Experten aufnehmen. 

Zurück zum Seitenanfang Zurück zum Seitenanfang