Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Baseline Noise

Empower Tip: Baseline Noise

Tip #298: Setting Up Baseline Noise

Tip #298: Setting Up Baseline Noise

In this tip for Empower Chromatography Data System (CDS), we will continue our series on Baseline Noise and discuss setting up Baseline Noise.

STEP 1

STEP 1

Setting up Baseline Noise involves identifying two regions over which noise will be calculated. Region 1 starts at the start time and ends at a distance in the positive direction equal to the Percent of Run Time to Average. Region 2 starts at the end time and goes the distance of the Percent of Run Time to Average in the negative direction. The area in the time range not incorporated in Region 1 or Region 2 is not used in the calculation. Region 1 and Region 2 can overlap if the start and end time are not set large enough.

STEP 2

STEP 2

In this example, the run time was 30 min. The noise region specified in our processing method was 26-28.3 min. The percent of run time to average was set at 7.5%. Therefore, the two noise regions created will be 2.25 min wide each. The Region 1 measured will be 26 to 28.25 min. Region 2 measured will be 26.05 to 28.3 min. Each region will be broken into 30-sec segments and an average for each region will be calculated. Then the two regions will be averaged to arrive at the final calculated amount of Baseline Noise.

STEP 3

STEP 3

Region 1 goes from 26 to 28.25 min. The time range is then broken up into 30-second segments. Notice that the last segment is not a full 30 seconds and will therefore not be used in the calculation. The highest point in each 30 sec. segment is subtracted from the lowest point in each 30 sec. segment to give the distance of noise for that segment. The segments are then averaged together to calculate the noise for Region 1.

STEP 4

STEP 4

The Region 2 segment is calculated in the same way as segment one except starting with 28.3 min. Thirty second segments are created, and noise is measured for each segment. The Segments are averaged to give a noise value for Region 2.

STEP 5

STEP 5

The reported Baseline Noise given by the Empower application is simply the average of Region 1 and Region 2. Note that although I’ve displayed them here in AU for consistencies sake, they are always reported in the application with units of mV.

It’s that easy!

Final Notes

For additional reference, view an example of Baseline Noise Tip 70

This can be done with either the Pro or QuickStart interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



Benötigen Sie Hilfe? Kontakt mit einem Waters Experten aufnehmen. 

Zurück zum Seitenanfang Zurück zum Seitenanfang