Empower Tip: Review Window and the Processing Method

Empower Tip: Review Window and the Processing Method

Tip #30: Calculating signal-to-noise ratio (Part 3)

Tip #30: Calculating signal-to-noise ratio (Part 3)

STEP 6

STEP 6

Now we need to identify the blank injection in the Sample Set. On the Sample Sets tab, right click on the Sample Set and select Alter Sample.

STEP 7

STEP 7

Right-click on the table and select Table Properties.

STEP 8

STEP 8

On the Columns tab, find the Blank field and deselect the check mark, then click OK.

STEP 9

STEP 9

Now that the Blank field is added to the table, check the box for the blank injection.

STEP 10

STEP 10

Save all changes and exit out of Alter Sample.

STEP 11

STEP 11

Process the data.

If you work in a pharmaceutical company you follow the pharmacopeia for your methodology. There are 2 ways Empower can calculate signal-to-noise ratio adhering to the pharmacopeia. This tip will address calculating noise from a blank injection.

STEP 1

STEP 1

Bring raw data (Channels, Injections or Sample Set) into Review and open the processing method. Click the Processing Parameters tool.

STEP 12

STEP 12

STEP 2

STEP 2

Bring the Result Set or an individual Result into Review.

Click the Suitability tab and check the box that says Calculate Suitability Results. Enter the Void Volume Time, an estimate is fine.

STEP 13

STEP 13

STEP 3

STEP 3

USP s/n, EP s/n, and JP s/n are hidden fields by default so let’s add one to our Peaks table. Right-click on the Peaks table and select Table Properties.

Select Calculate USP, EP, and JP s/n.

STEP 4

STEP 4

STEP 14

STEP 14

Select 'Use noise centered on peak region in blank injection'. Empower calculates peak-to-peak noise from the expected retention time for the peak(s) in the blank injection. The half-height multipliers set the window of time empower uses to calculate noise in the blank injection. For example, empower calculates the half-height width of a peak and multiplies by the value in the half-height multiplier field. If the peak width was 15 seconds at half height and the multiplier is 5, then the window would be 75 seconds. It does this for each peak.

On the Columns tab, find USP s/n and remove the check mark to make it visible in the table. Click OK.

STEP 5

STEP 5

STEP 15

STEP 15

Save all changes to the Processing Method and return to Browse Project.

The USP signal-to-noise ratio is displayed for every integrated peak.

STEP 16

STEP 16

To keep the USP s/n field in the peaks table, select Save Preferences from the File menu.

It’s that easy!

Final Notes

1. You must have the System Suitability option and enable it in the project.

2. This procedure can be followed for both the QuickStart and Pro Interfaces.