Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Featured Empower User Question

Empower Tip: Featured Empower User Question

Tip #313: Featured Empower User Question

Tip #313: Featured Empower User Question

This week, I would like to feature a recent question I received.

QUESTION:

We are running a method collecting two wavelengths per injection from a PDA and we want Empower to identify the channel with the higher peak response and report the higher value. How is that done?

For this example, I selected data and derived 240nm and 280nm from the 3D PDA Channel.

STEP 1

STEP 1

Create a simple peak, intersample Custom Field to calculate the ratio of the responses between the peaks in channel 1 to channel 2.

STEP 2:

Create a simple Custom Peak field to report the channel with the higher response.

STEP 3

STEP 3

Finally, create a Custom Peak field to report the larger response. Here is the full equation which is cut off in the screen capture:

LT(Reposne_Ratio,1)*(SAME.1.280nm(Area))+GT(Reposne_Ratio,1)*(SAME.1.240nm(Area))

STEP 4

STEP 4

An example result is provided.

It’s that easy!

Final Notes

This can be done with either the Pro or QuickStart interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



Benötigen Sie Hilfe? Kontakt mit einem Waters Experten aufnehmen. 

Zurück zum Seitenanfang Zurück zum Seitenanfang