Question: Why is the retention time of my peak slightly different from the retention time of the UV apex spectrum for the peak?
Great question and there is a simple explanation.
Once I extract a wavelength from a 3D PDA channel of data and integrate the chromatogram I see the retention times for the peaks are different from the retention times for the spectra extracted at their apices.
Peak retention time is determined by fitting a quadratic curve to the highest point in the peak and 2 points on either side of that highest point, and then determining the inflection point of that quadratic curve. The reason this is done is because if the peak is misshapen and you have collected data at a low sampling rate (set in the instrument method) the true apex of the peak may be between 2 actual data points. The retention time of the apex spectrum is simply taken from the highest point in the peak.
Follow-up Question: Can I set up Empower so that the retention time of the peak and the apex spectrum match?
The answer is YES provided you use the Traditional peak detection and integration algorithm. There is a system policy called Use v2.XX Style Retention Time Calculations, found on the Data Processing Tab of the System Policies. If enabled Empower will determine peak retention time as the highest point in the peak which is what was done in Millennium v2.XX.
We see the same chromatogram after the system policy has been enabled and the peak retention times match the apex spectra retention times.
It’s that easy!
1. The change in system policy only works for the Traditional peak detection and integration algorithm. The Apex Track algorithm is not affected by this system policy.
2. This procedure can be followed using the QuickStart or Pro interface.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
Benötigen Sie Hilfe? Kontakt mit einem Waters Experten aufnehmen.