.
Please refer back to Tip 87 which covered how to modify an existing Custom Field.
Let’s first review the result from Tip 88. The average peak width was calculated based on all peaks in the chromatogram which includes Found and Unknown peaks.
Let’s modify the Custom Field to calculate average peak width for only the Found peaks. Go to the properties of the project, Custom Fields tab, Highlight the Custom Field and click ‘Edit’. (Note: the Sample Type and Peak Type fields both default to All).
Click and view the choices in the drop down list for Sample Type. We could have this calculation done for only standards or only controls.
Click and view the choices in the drop down list for Peak Type. In this case we select Founds so that Empower makes the calculation for only Found peaks.
Once you make the choice, a warning message appears telling you that the Custom Field will be calculated for only the selected peak type. Click ‘OK’.
Once we reprocess our data, we see the average peak width value is different now that we have excluded the Unknown peak from the calculation.
It’s that easy!
This procedure can be followed using the QuickStart or Pro interface.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
Benötigen Sie Hilfe? Kontakt mit einem Waters Experten aufnehmen.