Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Review Window and the Processing Method

Empower Tip: Review Window and the Processing Method

Tip #99: When to batch process the sample set to get results from Custom Fields

Tip #99: When to batch process the sample set to get results from Custom Fields

Q: When do you need batch processing for Custom Fields? And can I generate results in Review with my Custom Fields?

A: Depends on the type of Custom Field.

In this tip, I will demonstrate when to batch process the sample set to get results from Custom Fields(Part 17).

STEP 1

STEP 1

If you recall from Tip 84 & Tip 85, we discussed how to calculate Percent Label Claim which is a relatively simple Peak Custom Field.

STEP 2

STEP 2

In this example, take the Sample Set into Review, open the Processing Method and review the samples one by one. The Percent Label Claim is calculated for the samples.

STEP 3

STEP 3

In last weeks’ Tip (Tip 98), we calculated the ‘average area’ of peaks over replicate injections of a sample. This is an Intersample Calculation with summary function ‘AVE’.

STEP 4

STEP 4

In this example, we need to ‘batch process’ the Sample Set with a Summarize Custom Field line at the end of the Sample Set so that the summary calculation works properly.

It’s that easy!

Final Notes

This procedure can be followed using the QuickStart or Pro interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



Benötigen Sie Hilfe? Kontakt mit einem Waters Experten aufnehmen. 

Zurück zum Seitenanfang Zurück zum Seitenanfang