Priming the system

Priming the system
Topic LCI-USG-0237, Last Updated 10/16/2026

After powering-on the system, you must prime it before the system is ready for use.

Requirement: You must prime the system after starting it, as well as after changing the mobile phase, after changing the sample needle, and after the system is idle for four hours or more.
Requirement: You must install a column before priming the system. See Installing the column.
Recommendation: If you are introducing new solvents, prime them at 4 mL/min for seven minutes. Alternatively, prime the solvents at 4 mL/min for three minutes. Ensure that sufficient quantities of solvent are available for priming.

There are multiple ways to prime the system.

Note: Starting with version 2.2.0, workflow progress indicators display the overall progress of the current workflow and the status of each workflow step. Steps are color-coded to show their current state: completed (green), in progress (amber), or pending (gray).

To prime the system:

  1. From the system touchscreen or the Console app, run the startup workflow to prime all solvents, prime the needle wash and seal wash, prime the sample metering pump, and specify the solvent composition, flow rate, and column and sample temperatures. See Running the startup workflow.
  2. To continue priming the system from the touchscreen (traditional mode) :
    1. Tap SETUP > Solvents > Prime Mobile Phase Solvents and follow the on-screen prompts to prime the pump.
      • Select which solvents to prime and priming duration
      • Set solvent composition and priming duration
      • Select equilibration settings
    2. Tap HOME, tap the solvent bottle icon, tap a mobile phase condition card, and then tap Prime Solvent to prime an individual mobile phase.
  3. To continue priming the system from the touchscreen (express mode) :
    1. To prime the pump, tap SETUP > Solvents, and then tap the play button next to Prime Mobile Phase Solvents. The summary page appears showing the last selected settings (or default settings upon first system startup).
      Optionally, tap CANCEL to exit the workflow before it begins, or click Back to adjust the workflow settings. After a 30-second countdown, the workflow runs automatically.
    2. Click DONE.
  4. To continue priming the system from the Console app (traditional mode) :
    1. Click Setup, click the left-hand side of the Prime Solvents Workflow card, click Start, and then follow the remaining on-screen prompts to complete the next steps. Actions include:
      • Select which solvents to prime and priming duration
      • Set solvent composition and priming duration
      • Select equilibration settings
    2. Click Done.
  5. To continue priming the system from the Console app (express mode) :
    1. To prime the pump, click Setup, and then click the play button on the right-hand side of the Prime Solvents Workflow card. The summary page appears showing the last selected settings (or default settings upon first system startup).
      Optionally, click Cancel to exit the workflow before it begins, or click Back to adjust the workflow settings. After a 30-second countdown, the workflow runs automatically.
    2. Click Done.