Confirming the health of the degasser

Confirming the health of the degasser
Topic LCI-USG-0233, Last Updated 11/21/2025

The degasser health check attempts to clear excess condensate from the degasser vent line, which can trigger a false alarm.

You can confirm the health of the degasser from the system touchscreen.

To confirm the health of the degasser:

  1. Tap HEALTH > Troubleshoot > Degasser Health Check > NEXT and follow the on-screen prompts.
  2. Tap START.
    The screen displays Degasser Health Check In Progress.
    Note: The degasser health check process takes approximately 60 minutes to complete.
  3. If desired, tap STOP to stop the degasser health check.
  4. The screen displays Degasser Health Check Passed. Tap DONE.