Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs! Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs!
Users can install optional UV filters in the PDA detector flow cell.
The Alliance iS PDA detector supports optional optical filter modules (214 nm long pass filter, 220 nm band pass filter, 260 nm band pass filter, or 280 nm band pass filter). The UV filters require an optional PDA flow cell that is specifically designed to be compatible with either a cuvette or the optional filters. For more information, see PDA detector UV filters (optional).
Caution: To avoid a Wavelength Verification Failure message, do not install the optional UV filter until after the PDA detector completes its startup sequence. The system verifies the PDA detector's calibration using two spectral peaks from the deuterium lamp and three absorbance peaks from its integrated erbium filter. Upon startup, the system verifies detector calibration by comparing the locations of these peaks with expected wavelengths based on calibration data stored in the detector's memory. If the results of this verification differ from the stored calibration by more than 1.0 nm, the detector displays a Wavelength Verification Failure message. If you install a UV filter before the system verifies the PDA detector's calibration, the verify calibration workflow fails. See
Calibrating the PDA detector and
Verifying the PDA detector's calibration for more information.
To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.
Refer to the English version of this page for the latest information.