Alliance iS HPLC system audit trail

Alliance iS HPLC system audit trail
Topic LCI-USG-0234, Last Updated 11/21/2025

The Alliance iS HPLC system tracks and records auditable user-performed activities in the system log and, optionally, in Empower.

When users with Empower version 3.8 or higher perform auditable activities from the touchscreen or the console, the system records those activities in the system log (see Viewing system logs for more information) and, optionally, in the Empower system audit trail (see Audit trail information in Empower and the Empower online Information System topic "Accessing the System Audit Trail"). The system records auditable, user-performed functions as tool kit in the action column of the audit trail table.

For users of Empower version 3 FR5 through Empower 3.7, auditable events are recorded in the system log (see Viewing system logs for more information) and the Empower message center. The system records auditable, user-performed functions as tool kit in the action column of the audit trail table. For details on the Empower message center, see the Empower online Information System topic "Accessing the message center".

Note: If the Empower Authentication feature is disabled, all user-performed activities are attributed to Unknown in the system logs.

Auditable events include:

  • Empower user sign in or out
  • Empower Authentication feature enabled or disabled
  • Create manual log entry
  • Admin
    • Change acquisition checks
    • Change qualification expiration date
    • Change PM expiration date
    • Run access control workflow
  • Replace lamp
  • Replace flow cell
  • Replace needle
  • Replace needle seal
  • Replace column
  • Calibrate detector
  • Verify calibration
  • Noise and drift
  • TUV autozero offsets
  • FTN
    • Sample needle metering pump leak test
    • Needle seal readiness
    • Sample compartment temp test
    • Change extension loop
    • Leak sensor enable/disable
    • Multi-draw valve enable/disable
    • Calibrate FTN motion axes (5x)
  • QSM
    • System leak test
    • Change mixer
    • Leak sensor
  • CHC
    • Compartment temp test
    • Leak sensor
    • eConnect reader
  • Degasser health check
  • Pump head maintenance
  • System
    • Change dwell volume
    • Change tubing kit
  • Detector leak sensors
  • Counters
    • Reset injection count
    • Reset liters pumped
  • Stop acquisition
  • Startup or shutdown workflow completed
  • Add event logs to the audit trail