Method Editor app Column Compartment view

Method Editor app Column Compartment view
Topic LCI-USG-0267, Last Updated 11/21/2025

When you select Column Compartment from the app's left-hand navigation panel, the page's middle panel populates with the options listed in the following table. Selecting any of these options populates the panel to the right with details or further options as described.

The app ensures that user-defined settings such as toggle selections and specified input values are retained across navigation events. This means that when you select options or specify values, those choices remain intact even when returning later after navigating away or closing the app. For any user-defined settings out of limits for the system, a red dot appears to the left of the middle panel option name.

Table 1. Alliance iS Method Editor app Column Compartment view options
Parameter Details or available settings provided
Column 1

The details for the selected column. To change the column type being used by the system, click the blue pencil icon (labeled Change in the panel's top right-hand corner) and, from the new view that opens, select the desired column type from the scrollable list. You can also search for a column type using the Search field. To exit the Change column view, click Cancel or Save.

Note: To ensure that the column type you specify is present on the system, enable the appropriate pre-run checks on that system.
Column Temperature

The Set Compartment Temperature option selector: Enable or disable control of the compartment temperature.

When enabled, the following items appear:

  • The Temperature Setpoint (°C) form field to specify a value, from 4.0 to 90.0.
  • The Set Temperature Tolerance option selector: Enable or disable to hold the next injections until the compartment temperature is within the specified tolerance. When enabled, the Tolerance (±°C) form field appears to specify a value, from ±1.0 to 10.0.
Data Channels: Column Compartment

The Column Temperature option selector: Enable or disable this data channel.

Note: The contents of this panel and available functionality are the same as when you select the Data Channels option from the navigation panel's System view.