Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Review Window and the Processing Method

Empower Tip: Review Window and the Processing Method

Tip #27: Using System Suitability to Flag Failing Results

Tip #27: Using System Suitability to Flag Failing Results

Wouldn’t it be nice if Empower could flag a failing result? Well, it can and it is very easy to set up.

STEP 1

STEP 1

Bring raw data (Channels, Injections or Sample Set) into Review and open the processing method. Click the Processing Method tool.

STEP 2

STEP 2

Click the Suitability tab and check the box that says Calculate Suitability Results. Enter the Void Volume Time, an estimate is fine.

STEP 3

STEP 3

Click the Limits tab. In the upper table (Suitability Components), highlight the row number for the component of interest.

STEP 4

STEP 4

In the middle table (Suitability Limits), click inside the Field Name field and choose the field of interest. This list contains all of the same peak fields you see in the Peaks table below the chromatogram in Review.

STEP 5

STEP 5

You can set a Target Value, however, that is not a requirement.

STEP 6

STEP 6

Enter a Lower Error Limit and an Upper Error Limit. Depending on the field of interest you may only need one limit. For example, I would only set a lower error limit for USP Plate Count. An upper error limit, in this case, wouldn’t apply because the higher the plate count, the better the system performance.

STEP 7

STEP 7

Save all changes to the Processing Method and exit from Review.

STEP 8

STEP 8

Back in project window, click the Sample Sets Tab, right-click on the Sample Set of interest, and select Process.

STEP 9

STEP 9

Select the Method Set or Processing Method, enter any required comments, and click OK.

STEP 10

STEP 10

Go to the Results tab and observe the Fault field.

STEP 11

STEP 11

Go to the Result Sets tab, right-click on the Result Set, and select Review.

STEP 12

STEP 12

As you drill down into the Result Set note any peak names in red, italicized font and then note any other fields in red, italicized font. This draws your attention to which components have issues and what those issues are.

STEP 13

STEP 13

Finally, if a limit for a peak field is exceeded, it will trigger Interactive System Suitability, which was covered in Tip 14 - Using Interactive System Suitability.

It’s that easy!

Final Notes

1. You must have the System Suitability option and enable it in the project.

2. You can set limits on a component-by-component basis.

2. This procedure can be followed for both the QuickStart and Pro Interfaces.



About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



Need Help? Connect with a Waters expert. 

Back To Top Back To Top