Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Empower User Question Answered

Empower Tip: Empower User Question Answered

Tip #377: Result Faults in Blanks for Must Peaks

Tip #377: Result Faults in Blanks for Must Peaks

In this tip, we will answer an Empower user question.

QUESTION:

Is it normal for my blank to be flagged when a Must Peak is absent?

Great question. Yes, it is normal behavior if a blank is flagged due to a missing Must Peak, Default Peak, RT Reference Peak, or Internal Standard Peak.

Let’s review.

STEP 1

STEP 1

This Sample Set has a blank, injected 3 times.

STEP 2

STEP 2

On the Component tab of the Processing Method, Peak A is selected as a Must Peak. This means that Peak A must be found in every chromatogram.

STEP 3

STEP 3

Peak A is not found in the blank, therefore, the result is flagged as Faulted and Result Code F01 is displayed.

STEP 4

STEP 4

A similar situation is observed if an Internal Standard peak, a Default peak, and/or an RT Reference peak is missing. Like a Must Peak, these types of peaks must also be present in every chromatogram. The online Help defines the Result Codes that indicate if one of these peaks is missing.

It’s that easy!



About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



Need Help? Connect with a Waters expert. 

Back To Top Back To Top