waters_connect 3.1.0 LTS Release Overview
Introduction
This page provides an overview of the waters_connect 3.1.0 LTS release, compatible waters_connect applications, instrument drivers and supported system configurations including liquid chromatography modules, mass spectrometry instruments, and detectors. For more information on waters_connect release types and associated level of software support please visit waters_connect Releases Overview.
Release Overview
Recommended Use
waters_connect helps you accelerate the time to outcomes with workflow solutions ranging from quantitative, biopharma, and accurate mass-based analyses helping you reach new levels of efficiency, precision, and confidence. Key solutions available with this release:
- waters_connect for Quantitation — Quantitative analysis with tandem quadrupole mass spectrometry systems.
- waters_connect for Biopharma — Biopharmaceutical accurate mass applications for protein-based therapeutics, vaccines, and nucleic acid-based therapeutics.
- waters_connect for Research & Routine — Bioprocess accurate mass applications for process (spent cell culture media) and product quality attribute monitoring including the Bioprocess Walk-Up Solution.
- waters_connect for Small Molecules — Accurate mass-based solutions for screening of small molecules.
Base Kit
Enables access to platform components and core applications such as Hub, Administration, System Console, Device Management, Explorer, Qualification for general system use.
Key benefits in this release include:
- increased lab operational security with support of Windows 10 LTSC 2021 for workstation, LND and Clients PCs.
- Improved data migration tool for a more efficient and secure data migration from UNIFI Software to waters_connect platform.
- Japanese localization for UNIFI application workflows with Xevo G2-XS QTof, Vion IMS QTof and RDa/BioAccord systems to elevate understanding of product functionalities.
- Improved usability with waters_connect Explorer application for managing records and accessing general properties and information recorded in the audit trail.
- Increased system uptime with platform in-place upgrade mechanism and simplified product deployment.
- Increased laboratory operations efficiency with support for up to 4 Xevo QTofs on a single network deployment & support for 512e disk drives for Server PCs.
LC-MS Toolkit
The LC-MS Toolkit application enables data interrogation outside of the defined workflows; provides flexibility to the software that is needed for more advanced use.
Sample Submission | Acquisition Method Editor
Enables access to acquisition utilities used within the Biopharmaceutical workflow with BioAccord & the MS Quan workflow with Xevo TQ-S cronos, Xevo TQ-S micro, Xevo TQ-XS and Xevo TQ Absolute.
Key benefits in this release include:
- Increased productivity with parallel acquisition and processing with Peptide MAM, Intact Mass and MS quan applications
- Increased Lab automation & throughput with the ability to integrate with the Sartorius Ambr 250 system
Peptide MAM
The Peptide Multi-Attribute Monitoring (MAM) application enables monitoring of biopharmaceutical product quality attributes (PQA) at the peptide level.
Key benefits in this release include:
- Increased usability and product functionality with the ability to download Scalable Vector Graphic (SVG) images; overlay theoretical isotope distributions over acquired data; use Extracted Ion Chromatogram (XIC) with most abundant isotope.
- Improved processing speed
Intact Mass
The Intact Mass application enables biologics Identity and Purity confirmation for mass confirmation and purity determination of oligonucleotides, peptides, protein and conjugates.
Key benefits in this release include:
- Reduced time results for process development studies with the support of the Sartorius Ambr 250 HT Data Integration with INTACT Mass App;
- Increase workflow automation with 3rd party application with the ability to fully automate Acquire/Process functionality for 3rd party informatics to gain controlled access to processed results generated by the INTACT Mass App.
- Extended workflow flexibility with the additional data region events for tuning chromatographic peak detection and processing, more flexibility in 5’-Cap modifiers for mRNA analysis, and additional predefined glycosylation modifiers for antibody/glycoprotein analysis.
CONFIRM Sequence
The CONFIRM Sequence application enables streamlined oligonucleotide sequence confirmation and impurity characterization workflow.
Key benefits in this release include:
- Improved quality and usability improvements to the Sequence Library, Analysis, Reporting, and Audit Trail functions of the application.
MS Quan
The MS Quan application enables routine small-molecule quantitation and accelerates quantitation data processing and review steps with exception-focused review and easy-to-view chromatogram windows for multiple samples.
Key benefits in this release include:
- New Standard Addition processing allowing quantitation from samples with complex matrix effects
- Performance Enhancements which reduce processing/reporting times and file sizes
- Copy the contents of the unknowns grid to the clipboard, for export to other applications for further processing
- Side panel updates to show more detailed information and version specific identifiers for improved traceability
Applications
Quantitative Analysis with Tandem Quadrupole Mass Spectrometry Systems
Acquisition
- Sample Submission 1.7.0
- Acquisition Method Editor 2.3.0
- Laboratory Network Device 1.11.0
Data View & Processing
- LC-MS Toolkit 1.7.0
- MS Quan 1.7.0
- MS Quan 1.9.0
MS Instrument (driver)
- Xevo TQ-S cronos (1.3.0)
- Xevo TQ-S micro (1.3.0)
- Xevo TQ-XS (1.3.0)
- Xevo TQ Absolute (1.3.0)
Biopharmaceutical Accurate Mass Applications for Protein-Based Therapeutics, Vaccines, and Nucleic Acid-Based Therapeutics
Acquisition
- UNIFI 3.1.0
- Laboratory Network Device 1.11.0
Data View & Processing
- LC-MS Toolkit 1.7.0
- LC-MS Toolkit 1.8.0
- UNIFI 3.1.0
- Confirm Sequence 1.2.0
- Confirm Sequence 1.2.1
MS Instrument (driver)
- ACQUITY RDa (3.1.0)
- Xevo G2-XS QTof (4.2.0)
- Xevo G3 QTof (1.1.0)
- Xevo G2-XS QTof (4.3.0)
- Xevo G3 QTof (1.2.0)
- Vion (4.2.0)
Acquisition
- Sample Submission 1.8.0
- Acquisition Method Editor 2.3.0
- Acquisition Method Editor 2.6.0
- Laboratory Network Device 1.11.0
Data View & Processing
- Peptide MAM 1.6.0
- Peptide MAM 1.5.0
- Intact Mass 1.4.0
- Intact Mass 1.3.0
- Confirm Sequence 1.2.1
MS Instrument (driver)
- BioAccord (3.1.0)
- Xevo G2-XS QTof (4.3.0)
- Xevo G3 QTof (1.2.0)
Acquisition
- Sample Submission 1.7.0
- Acquisition Method Editor 2.3.0
- Laboratory Network Device 1.11.0
Data View & Processing
- Peptide MAM 1.5.0
- Intact Mass 1.3.0
- Confirm Sequence 1.2.0
MS Instrument (driver)
- BioAccord (3.1.0)
Bioprocess Accurate Mass Applications for Process (Spent Cell Culture Media) and Product Quality Attribute Monitoring Including the Bioprocess Walk-Up Solution
Acquisition
- Sample Submission 1.8.0
- Acquisition Method Editor 2.3.0
- Laboratory Network Device 1.11.0
Data View & Processing
- UNIFI 3.1.0
- Peptide MAM 1.6.0
- Intact Mass 1.4.0
- Confirm Sequence 1.2.1
- LC-MS Toolkit 1.8.0
MS Instrument (driver)
- ACQUITY RDa (3.1.0)
Accurate Mass-Based Solutions for Screening of Small Molecules
Acquisition
- UNIFI 3.1.0
- Laboratory Network Device 1.11.0
Data View & Processing
- UNIFI 3.1.0
MS Instrument (driver)
- ACQUITY RDa (3.1.0)
- Xevo G2-XS QTof (4.2.0)
- Xevo G3 QTof (1.1.0)
- Xevo G2-XS QTof (4.3.0)
- Xevo G3 QTof (1.2.0)
- Vion (4.2.0)
Configurations
Operating System Support for Workstation Installs
Operating System
Windows
Requirements
- Windows Defender must be set to Off
- OS language: English (US), Japanese (Japan)
- Peptide MAM, Intact Mass, Confirm Sequence, Sample Submission, and LC-MS Toolkit support English (US) only
Versions Supported
- Windows 10 Enterprise LTSC 2021 edition (version 21H2)
- Windows 10 Pro version 22H2:
Operating System Support for Network Installs
Operating System
Windows
Versions Supported
Server PC
- Windows Server 2019 Standard
Requirements
- Windows Defender must be set to Off
- OS languages: English (US), Japanese (Japan)
- Peptide MAM, Intact Mass, Confirm Sequence, Sample Submission, and LC-MS Toolkit support English (US) only
LND PC
- Windows 10 Enterprise LTSC 2021 edition (version 21H2)
- Windows 10 Pro version 22H2
Requirements
- OS languages: English (US), Japanese (Japan)
Client PC
- Windows 10 Enterprise LTSC 2021 edition (version 21H2)
- Windows 10 Pro version 22H2
Requirements
- OS languages: Deployment supported on English and Japanese language versions of Windows. Hub is displayed in English (US) or Japanese (Japan) only.
- Peptide MAM, Intact Mass, Confirm Sequence, Sample Submission, and LC-MS Toolkit support English (US) only.
Mass spectrometry instruments supported on a single waters_connect network deployment:
|
|
|
|---|---|
Instrument |
Instrument |
Instrument |
Total Number of Systems |
|
ACQUITY RDa (Tof) |
ACQUITY RDa (Tof) |
|
ACQUITY RDa (Tof) |
up to 6 |
|
BioAccord LC-MS System (Tof) |
BioAccord LC-MS System (Tof) |
|
BioAccord LC-MS System (Tof) |
up to 6 |
|
Xevo G2-XS QTof |
Xevo G2-XS QTof |
|
Xevo G2-XS QTof |
up to 4 |
|
Xevo G3 QTof |
Xevo G3 QTof |
|
Xevo G3 QTof |
up to 4 |
|
Vion IMS QTof |
Vion IMS QTof |
|
Vion IMS QTof |
up to 2 |
|
Mix of Tof and QTof |
Mix of Tof and QTof |
|
Mix of Tof and QTof |
up to 4 (up to 2 Tof and up to 2 QTof) |
|
Mix of Tof |
Mix of Tof |
|
Mix of Tof |
up to 6 |
|
Mix of QTof |
Mix of QTof |
|
Mix of QTof |
up to 4 |
|
Quad |
Quad |
|
Quad |
not available |
ACQUITY Instrument Systems (such as pumps, detectors, sample managers, column compartments) supported by waters_connect driver pack (DP) 2022 R1
Analytical ACQUITY UPLC Systems
ACQUITY Premier/ACQUITY H-Class Binary Series (15,000 PSI limited systems)
UPLC Configuration
- Binary Solvent Manager (BSM)
- Sample Manager — Flow Through Needle (SM-FTN)
- Column Heater (CH-A)
- Column Manager (CM-A/CM-Aux) (with S up to 2 CM-Aux units using 6 columns)
Acquisition Apps
- Sample Submission (S)
- UNIFI (U)
Supported MS Instruments
- BioAccord (S & U)
- ACQUITY RDa (U)
- Xevo G2-XS QTof (S & U)
- Xevo G3 QTof (S & U)
ACQUITY Premier/ACQUITY H-Class Quaternary Series (15,000 PSI limited systems)
UPLC Configuration
- Quaternary Solvent Manager (QSM)
Acquisition Apps
- Sample Submission (S)
- UNIFI (U)
Supported MS Instruments
- BioAccord (S & U)
- ACQUITY RDa (U)
- Xevo G2-XS QTof (S & U)
- Xevo G3 QTof (S & U)
UPLC Configuration
- Sample Manager — Flow Through Needle (SM-FTN)
Acquisition Apps
- UNIFI
Supported MS Instruments
- BioAccord
- ACQUITY RDa
- Xevo G2-XS QTof
- Xevo G3 QTof
UPLC Configuration
- Column Heater (CH-A)
Acquisition Apps
- Sample Submission (S)
- UNIFI (U)
Supported MS Instruments
- BioAccord (S & U)
- ACQUITY RDa (U)
- Xevo G2-XS QTof (S & U)
- Xevo G3 QTof (S & U)
UPLC Configuration
- Column Manager (CM-A/CM-Aux)
Acquisition Apps
- UNIFI
Supported MS Instruments
- BioAccord
- ACQUITY RDa
- Xevo G2-XS QTof
- Xevo G3 QTof
ACQUITY UPLC I-Class Plus Series (18,000 PSI limited systems)
UPLC Configuration
- Solvent Manager (BSM)
- Sample Manager — Flow Through Needle (SM-FTN)
Acquisition Apps
- UNIFI
Supported MS Instruments
- BioAccord (S & U)
- ACQUITY RDa (U)
- Xevo G2-XS QTof (S & U)
- Xevo G3 QTof (S & U)
UPLC Configuration
- Sample Manager — Fixed Loop (SM-FL)
Acquisition Apps
- UNIFI
Supported MS Instruments
- BioAccord
- ACQUITY RDa
- Xevo G2-XS QTof
- Xevo G3 QTof
UPLC Configuration
- Column Heater (CH-A)
- Column Manager (CM-A)
Acquisition Apps
- Sample Submission (S)
- UNIFI (U)
Supported MS Instruments
- BioAccord (S & U)
- ACQUITY RDa (U)
- Xevo G2-XS QTof (S & U)
- Xevo G3 QTof (S & U)
ACQUITY UPLC Series (15,000 PSI limited systems)
UPLC Configuration
- Binary Solvent Manager (BSM)
- Sample Manager — Fixed Loop (SM-XY)
- Column Heater HTCH
- Column Manager CM/CHC
Acquisition Apps
- UNIFI
Supported MS Instruments
- ACQUITY RDa
- Xevo G2-XS QTof
Note: ACQUITY modules firmware versions are detailed in the ACQUITY UPLC drive pack release notes.
ACQUITY Detectors
ACQUITY Detectors
UPLC Configuration
- ACQUITY UPLC Fluorescence (FLR) Detector
Acquisition Apps
- UNIFI
Supported MS Instruments
- BioAccord
- ACQUITY RDa
- Xevo G2-XS QTof
- Xevo G3 QTof
UPLC Configuration
- ACQUITY UPLC Tunable UV (TUV) Detector
- ACQUITY UPLC Photodiode Array (PDA)/eLambda PDA Detector (1; 2)
Acquisition Apps
- Sample Submission (S)
- UNIFI (U)
Supported MS Instruments
- BioAccord (S & U)
- ACQUITY RDa (U)
- Xevo G2-XS QTof (S & U)
- Xevo G3 QTof (S & U)
Note: ACQUITY modules firmware versions are detailed in the ACQUITY UPLC drive pack release notes.
ACQUITY Add-On Modules
ACQUITY Modules
UPLC Configuration
- ACQUITY Common Platform Sample Organizer (CPSO)
Acquisition Apps
- Sample Submission (S)
- UNIFI (U)
Supported MS Instruments
- BioAccord (S & U)
- ACQUITY RDa (U)
- Xevo G2-XS QTof (S & U)
- Xevo G3 QTof (S & U)
UPLC Configuration
- eSAT/IN
Acquisition Apps
- UNIFI
Supported MS Instruments
- BioAccord
- ACQUITY RDa
- Xevo G2-XS QTof
- Xevo G3 QTof
Note: ACQUITY modules firmware versions are detailed in the ACQUITY UPLC drive pack release notes.