waters_connect 3.3.0 Release Overview

waters_connect 3.3.0 Release Overview

Introduction

Introduction

This page provides an overview of the waters_connect 3.3.0 release, compatible waters_connect applications, instrument drivers and supported system configurations including liquid chromatography modules, mass spectrometry instruments, and detectors. For more information on waters_connect release types and associated level of software support please visit waters_connect Releases Overview.

Release Overview

Recommended Use

waters_connect helps you accelerate the time to outcomes with workflow solutions ranging from quantitative, biopharma, and accurate mass-based analyses helping you reach new levels of efficiency, precision, and confidence.

Key Solutions Available with this Release

Key Benefits

Key Benefits

Base Kit

Base Kit

waters_connect Base Kit comprises platform components and core applications such as Hub, Administration, UNIFI, System Console, Device Management, Explorer, and Qualification for general system use.

Key benefits in this release include:

  • Improved data clarity to achieve faster decision making when using the UNIFI Accurate Mass screening workflow with Select Series MRT data. Benefit from the enhanced isotope clustering algorithm improvements and leverage  higher spectral resolution data for species containing N15, S34 and O18, and C13 and Cl35;

  • Improved product quality and stability by addressing high impact software anomalies, such as:

    • inability to import library components with detection result having adducts and charge states; 

    • analyte names not showing in the channel filter of MRM MassLynx data imported into waters_connect;

    • slow performance when discarding a batch with a large number of records; 

    • Masslynx data import failures when using a network share;

    • lack of information on root cause and issue resolution when processing completes with errors in the UNIFI application; 

    • import MassLynx data limitation not showing the Quan Reference value.

Acquisition Kit

Acquisition Kit

Sample Submission and Method Editor enable access to acquisition utilities used for accurate mass Biopharmaceutical applications with Xevo G2-XS QTof, Xevo G3 QTof, and BioAccord System, and the quantitation workflow with Xevo TQ-S cronos, Xevo TQ-S micro, Xevo TQ-XS, and Xevo TQ Absolute.  

Key benefits in this release include:

  • Increased lab productivity and throughput with the introduction of support for the ACQUITY PDA Detector and ACQUITY Sample Organizer modules within the Sample Submission and Method Editor applications;

  • Increased usability and productivity for accurate mass workflows with the support of Data Dependent Acquisitions (DDA) within the Sample Submission and Method Editor applications when using Xevo G2-XS QTof and Xevo G3 QTof systems;

  • Increased usability and reduced waiting time for system readiness with new default instrument systems related settings, such as: 

    • view by default the Instrument System details panel when accessing the System Console application;

    • new defaults parameter for source temperature and scan time in the Method Editor application when using the Xevo G2-XS QTof and Xevo G3 QTof systems;

LC-MS Toolkit

LC-MS Toolkit

LC-MS Toolkit enables data interrogation outside of the defined workflows; provides flexibility to the software that is needed for more advanced use.

Key benefits in this release include:

  • Increased data clarity and faster decision making by viewing and manipulating 2D PDA in the LC-MS Toolkit application (such as 2D channel, PDA Max plot, deuterium spectra, extract wavelength spectra, extract wavelength chromatogram);

  • Faster time to results of unknown components with enhanced structural elucidation capabilities, such as:

    • Mass Fragment for determining the likely fragments from a mass spectrum from a purposed .mol file;

    • Elemental Composition improvements, including the ability to load and save user settings and to select different charge states for the precursor and product ions.

Peptide MAM

Peptide MAM

Peptide Multi-Attribute Method (MAM) enables monitoring of biopharmaceutical product quality attributes (PQA) at the peptide level.

Key benefits in Peptide MAM 1.8.2 include:

  • General maintenance release
  • Defect fixes

Intact Mass

Intact Mass

Intact Mass enables biologics mass confirmation and purity determination of oligonucleotides, peptides, protein and conjugates.

Key benefits in this release include:

  • Automated Calculations of DAR and Drug Distribution for Conjugated Molecule Analysis - Users can now deploy a BioAccord System, Xevo G3 QTof, or Xevo MRT QTof to analyze conjugated molecules, specifying the conjugate and its mass in the INTACT Mass data processing parameters, and have the calculation of Drug Antibody Ratio and distribution of conjugate forms automatically performed from the acquired data.

  • Controls for manual adjustments of chromatographic integration - Users can now manually adjust chromatographic peak integration and associated peak identification results and use those reintegrated peaks to direct the processing of mass spectral data.

  • Sample List Entry of Chemical Formula - User can now specify a ‘Chemical formula for deconvolution’ that will be used to refine the deconvolution isotope model, improving deconvolution mass accuracy and quantification.

  • Updated Sample List Column ‘Expected Masses’ changed to ‘Expected Masses and/or Chemical Formulas’: Users now can enter targets as numeric masses, as chemical formulas, or a mixture of both.

  • New Method option within ‘Specify calculation parameters’, allowing the user more control over the ‘Main Peak Retention Time Window’ to improve main peak identification in Optical Only and Variant Analysis workflows

  • Bioprocessing workflow: Additional sample list columns for Time Point and Dilution Factor that facilitates processing time series trending data (in bioprocess monitor or in third party solutions) and dilution factor to enable clearer results visualization and trending.

Synthetic Library

Synthetic Library

Synthetic Library supports the Confirm Sequence and Map Sequence applications for oligo sequencing and mapping workflows.

Key Benefits in Synthetic Library 2.0 include:

  • Support for oligo mapping workflow - new enzymes and polynucleotides sections

  • Multiple library support including user access controls

CONFIRM Sequence

CONFIRM Sequence

CONFIRM Sequence enables streamlined oligonucleotide sequence confirmation and impurity characterization workflow

Key benefits in CONFIRM Sequence 2.0 include:

  • Compatibility with Synthetic Library 2.0.0
  • Maximum charge state support extended to -20
  • The user can now confirm if their chosen nucleic acid digestion strategies can theoretically produce the desired sequence coverage, using a visualization of theoretical digest sequence coverage, including calling out segments of isobaric mass ambiguity. This helps users to determine the best digestion strategy to achieve optimal sequence coverage using unique digestion products.
  • Localization into Japanese

 

Map Sequence

Map Sequence

Map Sequence enables accurate mass mapping of digested oligonucleotides.

Key Benefits in Map Sequence 2.0 include:

  • Supported by Synthetic Library for sequence entry and experimental planning

  • Oligo mapping and sequence confirmation using LCMSE precursor and fragment data

  • Interactive annotated chromatogram

  • Interactive coverage map for simple visualization of sequence coverage and navigation to supporting data

  • Drill-down to fragment analysis

  • Dynamic pass-fail filtering and acceptance criteria for investigation and curation of results

  • Integrated reporting

MS Quan

MS Quan

MS Quan enables routine small-molecule quantitation and accelerates quantitation data processing and review steps with exception-focused review and easy-to-view chromatogram windows for multiple samples.

Key benefits in this release include:

  • Explicit edit and discard actions on methods and results to clarify the situation around saving of data;
  • clear display of version information for all MS Quan entities;
  • automatic calculation of a recovery % for Primary Internal Standards;
  • Reporting enhancements to allow users to further define the content of the PDF report, such as:
    • Option to set the size of chromatograms and define the axes
    • Filtering of injections/analytes included
    • Inclusion of instrument system type and name
  • Improvements to the manual integration workflows that allow new peaks to be added and defined more reliably.

 

Note: full information on new features and quality improvements available in the release notes.

Applications

Quantitative Analysis with Tandem Quadrupole Mass Spectrometry Systems

Quantitative Analysis with Tandem Quadrupole Mass Spectrometry Systems

Acquisition

  • Sample Submission 2.3.0 [new] 
  • Sample Submission 2.2.1
  • Acquisition Method Editor 3.3.0 [new] 
  • Laboratory Network Device 1.16.1

 

Data View & Processing

  • LC-MS Toolkit 1.13.0
  • MS Quan 1.11.0
  • MS Quan 1.11.1 
  • MS Quan 1.12.0 [new]

Instrument Drivers

  • Xevo TQ-S cronos 1.4.0
  • Xevo TQ-S micro 1.4.0
  • Xevo TQ-XS 1.4.0
  • Xevo TQ Absolute 1.4.0
  • Xevo TQ-S cronos 1.6.0
  • Xevo TQ-S micro 1.6.0
  • Xevo TQ-XS 1.6.0
  • Xevo TQ Absolute 1.6.0
  • ACQUITY UPLC DP 2024 R1 3.6.0
  • ACQUITY UPLC DP 3.7.0 [new]

Biopharmaceutical Accurate Mass Applications for Protein-Based Therapeutics, Vaccines, and Nucleic Acid-Based Therapeutics

Biopharmaceutical Accurate Mass Applications for Protein-Based Therapeutics, Vaccines, and Nucleic Acid-Based Therapeutics

Acquisition

  • UNIFI 3.8.0 [new] 
  • Laboratory Network Device 1.16.1

Data View & Processing

  • LC-MS Toolkit 1.13.0 
  • UNIFI 3.8.0
  • Synthetic Library 2.0.0
  • Confirm Sequence 2.0.0
  • Map Sequence 2.0.0

Instrument Drivers

  • ACQUITY RDa 3.3.0
  • Xevo G2-XS QTof 4.4.0
  • Xevo G3 QTof 1.3.0
  • Vion IMS QTof 4.3.0
  • ACQUITY UPLC DP 2024 R1 3.6.0
  • ACQUITY UPLC DP 3.7.0 [new]

Acquisition

  • Sample Submission 2.3.0 [new]
  • Sample Submission 2.2.1
  • Acquisition Method Editor 3.3.0 [new] 
  • Laboratory Network Device 1.16.1

Data View & Processing

  • Peptide MAM 1.8.2
  • Intact Mass 1.9.0
  • Synthetic Library 2.0.0
  • Confirm Sequence 2.0.0
  • Map Sequence 2.0.0

Instrument Drivers

  • BioAccord 3.3.0
  • Xevo G2-XS QTof 4.4.0
  • Xevo G3 QTof 1.3.0
  • Vion (not supported)
  • ACQUITY UPLC DP 2024 R1 3.6.0
  • ACQUITY UPLC DP 3.7.0 [new]

Bioprocess Accurate Mass Applications for Process (Spent Cell Culture Media) and Product Quality Attribute Monitoring Including the Bioprocess Walk-Up Solution

Bioprocess Accurate Mass Applications for Process (Spent Cell Culture Media) and Product Quality Attribute Monitoring Including the Bioprocess Walk-Up Solution

Acquisition

  • Sample Submission 2.3.0 [new] 
  • Sample Submission 2.2.1 
  • Acquisition Method Editor 3.3.0 
  • Laboratory Network Device 1.16.1

Data View & Processing

  • LC-MS Toolkit 1.13.0 
  • UNIFI 3.8.0
  • Intact Mass 1.9.0

Instrument Drivers

  • ACQUITY RDa 3.3.0 only

Accurate Mass-Based Solutions for Screening of Small Molecules

Accurate Mass-Based Solutions for Screening of Small Molecules

Acquisition

  • UNIFI 3.8.0 
  • Laboratory Network Device 1.16.1

Data View & Processing

  • UNIFI 3.8.0

Instrument Drivers

  • ACQUITY RDa 3.3.0
  • Xevo G2-XS QTof 4.4.0
  • Xevo G3 QTof 1.3.0
  • Vion (not supported)
  • Vion IMS QTof 4.3.0
  • ACQUITY UPLC DP 2024 R1 3.6.0
  • ACQUITY UPLC DP 3.7.0 [new]

[new] = latest application release.

Configurations

Operating System Support for Workstation Installs

Operating System Support for Workstation Installs

Operating System
Windows

Requirements

  • Windows Defender must be set to Off.
  • OS language: English (US), Japnese (Japan)
  • Windows images provided with waters_connect PCs include Windows IoT Enterprise licensing which enables extended support on the operating system end of service over the standard edition. View more information on Windows releases.

Versions Supported
Workstation PC

  • Windows 10 Enterprise IoT LTSC 2021 edition (version 21H2)
  • Windows 10 Pro version 22H2

Operating System Support for Network Installs

Operating System Support for Network Installs

Operating System
Windows

Versions Supported

Server PC

  • Windows Server 2019 Standard

Requirements

  • Windows Defender must be set to Off
  • OS languages: English (US), Japanese (Japan)

LND PC

  • Windows 10 Enterprise IoT LTSC 2021 edition (version 21H2)
  • Windows 10 Pro version 22H2

Requirements

  • OS languages: English (US), Japanese (Japan)
  • Windows images provided with waters_connect PCs include Windows IoT Enterprise licensing which enables extended support on the operating system end of service over the standard edition. View more information on Windows releases.

Client PC

  • Windows 10 Enterprise IoT LTSC 2021 edition (version 21H2)
  • Windows 10 Pro version 22H2

Requirements

  • OS languages: Deployment possible on European and Japanese language versions of Windows. Hub is displayed in English (US) or Japanese (Japan) only.

Mass spectrometry instruments supported on a single waters_connect network deployment:

Mass spectrometry instruments supported on a single waters_connect network deployment:

Instrument Type

Instrument Type

Instrument Type

Instrument Type

Instrument Type

Instrument Type

Instrument

Total Number of Systems

Recommended Number of Client PCs

Recommended Number of Users

Tof

Tof

Tof

Tof

Tof

Tof

  • ACQUITY RDa
  • BioAccord LCMS system
  • Mix of Tof

up to 6

up to 6

  • up to 6 (concurrent)
  • up to 25 (active software accounts)

QTof

QTof

QTof

QTof

QTof

QTof

  • Xevo G2-XS QTof
  • Xevo G3 QTof
  • Vion IMS QTof
  • Mix of QTof

up to 4

up to 4

  • up to 4 (concurrent)
  • up to 25 (active software accounts)

Tof & QTof

Tof & QTof

Tof & QTof

Tof & QTof

Tof & QTof

Tof & QTof

  • Mix of Tof and QTof

up to 4

up to 4

  • up to 4 (concurrent)
  • up to 25 (active software accounts)

TQ

TQ

TQ

TQ

TQ

TQ

  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo TQ Absolute
  • Mix of TQ

up to 5        

up to 5

  • up to 10 (concurrent)
  • up to 25 (active software accounts)

TQ, Tof, & QTof

TQ, Tof, & QTof

TQ, Tof, & QTof

TQ, Tof, & QTof

TQ, Tof, & QTof

TQ, Tof, & QTof

  • Mix of Quads
  • QTofs/Tofs

up to 6 in the specified configurations:

  • up to 3 TQ + up to 3 QTof
  • up to 4 TQ + up to 2 QTof
  • up to 3 TQ + up to 2 QTof + 1 Tof
  • up to 3 TQ + 1 QTof + 2 Tof
  • up to 3 TQ + up to 3 Tof

up to 6

  • up to 6 (concurrent)
  • up to 25 (active software accounts)

ACQUITY Instrument Systems (such as pumps, detectors, sample managers, column compartments) supported by waters_connect driver packs ACQUITY UPLC DP 2024 R1 v3.6.0 &  ACQUITY UPLC DP v3.7.0

ACQUITY Instrument Systems (such as pumps, detectors, sample managers, column compartments) supported by waters_connect driver packs ACQUITY UPLC DP 2024 R1 v3.6.0 &  ACQUITY UPLC DP v3.7.0

Microscale ACQUITY UPLC Systems

Microscale ACQUITY UPLC Systems

ACQUITY UPLC M-Class

ACQUITY UPLC M-Class

UPLC Configuration

  • uBinary Solvent Manager
  • uSample Manager-Fixed Loop
  • Auxillary Solvent Manager
  • Trap Valve Manager

Acquisition Apps

  • Sample Submission

Supported MS Instruments

  • Xevo G3 QTof with NanoLockSpray and low-flow ESI probe
  • Xevo G2-XS QTof with low-flow ESI probe only 

Note: ACQUITY modules firmware versions are detailed in the ACQUITY UPLC drive pack release notes.

Analytical ACQUITY UPLC Systems

Analytical ACQUITY UPLC Systems

ACQUITY Premier/ACQUITY H-Class Binary Series (15,000 PSI limited systems)

ACQUITY Premier/ACQUITY H-Class Binary Series (15,000 PSI limited systems)

UPLC Configuration

  • Binary Solvent Manager (BSM)
  • Sample Manager - Flow Through Needle (SM-FTN)
  • Column Heater (CH-A)
  • Column Manager (CM-A/CM-Aux) (with S up to 2 CM-Aux units using 6 columns)

Acquisition Apps

  • Sample Submission (S)
  • UNIFI (U)

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)
  • BioAccord (S & U)
  • ACQUITY RDa (U)
  • Xevo G2-XS QTof (S & U)
  • Xevo G3 QTof (S & U)
  • Vion IMS QTof (U)

ACQUITY Premier/ACQUITY H-Class Quaternary Series (15,000 PSI limited systems)

ACQUITY Premier/ACQUITY H-Class Quaternary Series (15,000 PSI limited systems)

UPLC Configuration

  • Quaternary Solvent Manager (QSM)

Acquisition Apps

  • Sample Submission (S)
  • UNIFI (U)

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)
  • BioAccord (S & U)
  • ACQUITY RDa (U)
  • Xevo G2-XS QTof (S & U)
  • Xevo G3 QTof (S & U)
  • Vion IMS QTof (U)

UPLC Configuration

  • Sample Manager - Flow Through Needle (SM-FTN)

Acquisition Apps

  • Sample Submission

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)

UPLC Configuration

  • Sample Manager - Flow Through Needle (SM-FTN)

Acquisition Apps

  • UNIFI

Supported MS Instruments

  • BioAccord
  • ACQUITY RDa
  • Xevo G2-XS QTof
  • Xevo G3 QTof
  • Vion IMS QTof

UPLC Configuration

  • Column Heater (CH-A)

Acquisition Apps

  • Sample Submission (S)
  • UNIFI (U)

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)
  • BioAccord (S & U)
  • ACQUITY RDa (U)
  • Xevo G2-XS QTof (S & U)
  • Xevo G3 QTof (S & U)
  • Vion IMS QTof (U)


UPLC Configuration

  • Column Manager (CM-A/CM-Aux) (with S up to 2 CM-Aux units using 6 columns)

Acquisition Apps

  • Sample Submission

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)

UPLC Configuration

  • Column Manager (CM-A/CM-Aux) (with S up to 2 CM-Aux units using 6 columns)

Acquisition Apps

  • UNIFI

Supported MS Instruments

  • BioAccord
  • ACQUITY RDa
  • Xevo G2-XS QTof
  • Xevo G3 QTof
  • Vion IMS QTof

ACQUITY UPLC I-Class Plus Series (18,000 PSI limited systems)

ACQUITY UPLC I-Class Plus Series (18,000 PSI limited systems)

UPLC Configuration

  • Solvent Manager (BSM)
  • Sample Manager - Flow Through Needle (SM-FTN)

Acquisition Apps

  • Sample Submission (S)
  • UNIFI (U)

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)
  • BioAccord (S & U)
  • ACQUITY RDa (U)
  • Xevo G2-XS QTof (S & U)
  • Xevo G3 QTof (S & U)

UPLC Configuration

  • Sample Manager – Fixed Loop (SM-FL)

Acquisition Apps

  • Sample Submission

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)

UPLC Configuration

  • Sample Manager – Fixed Loop (SM-FL)

Acquisition Apps

  • UNIFI

Supported MS Instruments

  • BioAccord
  • ACQUITY RDa
  • Xevo G2-XS QTof
  • Xevo G3 QTof
  • Vion IMS QTof

UPLC Configuration

  • Column Heater (CH-A)
  • Column Manager (CM-A)

Acquisition Apps

  • Sample Submission (S)
  • UNIFI (U)

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)
  • BioAccord (S & U)
  • ACQUITY RDa (U)
  • Xevo G2-XS QTof (S & U)
  • Xevo G3 QTof (S & U)
  • Vion IMS QTof (U)

ACQUITY UPLC Series (15,000 PSI limited systems)

ACQUITY UPLC Series (15,000 PSI limited systems)

UPLC Configuration

  • Binary Solvent Manager (BSM)
  • Sample Manager – Fixed Loop (SM-XY)
  • Column Heater HTCH
  • Column Manager CM/CHC

Acquisition Apps

  • UNIFI

Supported MS Instruments

  • ACQUITY RDa
  • Xevo G2-XS QTof

Note: ACQUITY modules firmware versions are detailed in the ACQUITY UPLC drive pack release notes.

ACQUITY Detectors

ACQUITY Detectors

ACQUITY Detectors

ACQUITY Detectors

UPLC Configuration

  • ACQUITY UPLC Fluorescence (FLR) Detector

Acquisition Apps

  • UNIFI

Supported MS Instruments

  • BioAccord
  • ACQUITY RDa
  • Xevo G2-XS QTof
  • Xevo G3 QTof
  • Vion IMS QTof

UPLC Configuration

  • ACQUITY UPLC Tunable UV (TUV) Detector
  • ACQUITY UPLC Photodiode Array (PDA) / eLambda PDA Detector (1; 2) 

Acquisition Apps

  • Sample Submission (S)
  • UNIFI (U)

Supported MS Instruments

  • Xevo TQ-S cronos (S)*
  • Xevo TQ-S micro (S)*
  • Xevo TQ-XS (S)*
  • Xevo TQ Absolute (S)*
  • BioAccord (S & U)
  • ACQUITY RDa (U)
  • Xevo G2-XS QTof (S & U)
  • Xevo G3 QTof (S & U)
  • Vion IMS QTof (U)

*Note: for the waters_connect for Quantitation workflow data interrogation / view available only with LC-MS Toolkit for Optical data. ACQUITY modules firmware versions are detailed in the ACQUITY UPLC drive pack release notes.

ACQUITY Add-On Modules

ACQUITY Add-On Modules

ACQUITY Modules

ACQUITY Modules

UPLC Configuration

  • ACQUITY Common Platform Sample Organizer (CPSO)

Acquisition Apps

  • Sample Submission (S)
  • UNIFI (U)

Supported MS Instruments

  • Xevo TQ-S cronos (S)
  • Xevo TQ-S micro (S)
  • Xevo TQ-XS (S)
  • Xevo TQ Absolute (S)
  • BioAccord (S & U)
  • ACQUITY RDa (U)
  • Xevo G2-XS QTof (S & U)
  • Xevo G3 QTof (S & U)
  • Vion IMS QTof (U)

UPLC Configuration

  • eSAT/IN

Acquisition Apps

  • UNIFI

Supported MS Instruments

  • BioAccord
  • ACQUITY RDa
  • Xevo G2-XS QTof
  • Xevo G3 QTof
  • Vion IMS QTof

Note: ACQUITY modules firmware versions are detailed in the ACQUITY UPLC drive pack release notes.

Supporting Files