Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Empower CDS User Question

Empower Tip: Empower CDS User Question

Tip #260: How to calculate resolution between peaks that are in different channels, using System Suitability

Tip #260: How to calculate resolution between peaks that are in different channels, using System Suitability

In this tip for Empower Chromatography Data System (CDS), we will answer an Empower CDS user question.

Q: How can I calculate the resolution between two peaks that are in two different channels?

Great question. There are two ways you can do this, and this tip addresses the first way (i.e., combining the two channels into one channel and then calculating the resolution between the two peaks. It doesn’t matter what type of data you are working with: UV, fluorescence, mass spec, and so forth). In this example, we are working with SIR channels from the QDa mass detector.

STEP 1

STEP 1

We have a stack plot of the two channels of data, each with its own peak eluting at different retention times. In this example, we can add the two channels together and then calculate the resolution between the peaks.

STEP 2:

Open the Method Set that was used to collect and process the data.

STEP 3

STEP 3

Right click on ‘Derived Channels’, select new ‘Derived Channel’ and on the ‘First (Only) Channel’ tab select the first channel from the drop-down list.

STEP 4

STEP 4

On the Second Channel tab, select ‘+’ from the ‘Operator’ drop down list. Select the second channel from the drop-down list. Check the ‘From injection Labeled’ box and enter the label for the injection.

STEP 5

STEP 5

Click ‘OK’ and name the derived channel. It now appears in the channel table along with the two original channels.

STEP 6

STEP 6

Apply the Method Set. You now see the combined chromatogram.

STEP 7

STEP 7

The resolution is now calculated between the two peaks.

It's that easy!

Final Notes

This can be done with either the Pro or QuickStart interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



Serve Aiuto? Contattate un esperto Waters. 

Torna all'inizio Torna all'inizio