In this tip for Empower Chromatography Data System (CDS), we continue our series on Calculating Amount/Concentration, Calibration Curve fit types, and review Linear Fits.
For a Linear Fit, Empower calculates the line that best fits the amount (or concentrations) versus the responses. The equation is Y = Bx + A where B is the slope and A is the Y intercept.
For an Inverse Linear Fit, Empower performs a Linear Fit to the X and Y values of the calibration points. The X value used in this case is 1/X for each standard peak. The equation used is the same as Linear Fit.
For a Log-Log Fit, Empower performs a Linear Fit to the X and Y values of the calibration points. The Y value of each point is the Log of the response of the peaks in the standard. The X value of each point is the Log of the amount or concentration of the peaks in the standard. The equation used is the same as Linear Fit.
For a Power Curve Fit, Empower performs a Linear Fit of log Y versus log X, where Y is the response and X is the X value.
The plot and the equation of the cure are reported in terms of the X value and response.
It’s that easy!
This can be done with either the Pro or QuickStart interface.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
Serve Aiuto? Contattate un esperto Waters.