In this tip for Empower Chromatography Data System (CDS), we continue our series on Calculating Amount/Concentration, Calibration Curve fit types, and review Quadratic Fit.
For a Quadratic Fit, Empower performs a least-squares fit of a quadratic polynomial to the calibration points. A minimum of 5 points is recommended.
The Quadratic Fit generates a Calibration Curve that is represented by the equation in figure 2.
Calculating amount or concentration in the sample is determined by solving for X in the equation. Empower only reports positive values of X that are within the range of the Calibration Curve.
Calibration Curves can be accessed in various ways, for example:
Review – bring a result into Review and click on the Calibration Curve tool
Compare – right click on a curve from the Curves tab in the project and select Compare
It’s that easy!
This can be done with either the Pro or QuickStart interface.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
Serve Aiuto? Contattate un esperto Waters.