Tip #308: Calculating USP Signal-to-Noise

Tip #308: Calculating USP Signal-to-Noise

In this tip for Empower Chromatography Data System (CDS), we will pause our series on peak identification to answer an Empower user question on calculating USP signal-to-noise.

Question: How do I calculate USP signal-to-noise using a specified time region in the blank for the noise?

This can be done with a peak Custom Field using intersample syntax.

STEP 1

STEP 1

USP signal-to-noise (S/N) is calculated as 2xHeight of peak/noise.

STEP 2

STEP 2

From the Sample Set Method, Blank is the label for the blank injection and there is only one injection. This is single-channel UV data, therefore, there is no need to specify the channel.

STEP 3

STEP 3

Create a peak Custom Field using intersample syntax. We need the Scale to µV field in the equation because peak height is in µV (microvolts) and peak-to-peak noise is in detector units which in this case is AU.

STEP 4

STEP 4

Go to the Suitability tab in the Processing Method and select peak-to-peak noise from the Noise Value for s/n drop-down.

STEP 5

STEP 5

Go to the Noise and Drift tab in the Processing Method and select the time region from the blank injection from which peak-to-peak noise will be calculated.

STEP 6

STEP 6

Once the Sample Set is processed, take the Result Set into Review. Go to the Results window and in the Chromatogram Result table, we can see the peak-to-peak noise calculated from the blank injection.

STEP 7

STEP 7

Click on a result for one of the standards or samples and the Custom Field is calculated based on the noise from the blank injection specified in the Custom Field.

Important Note: peak-to-peak noise will be calculated for every injection. However, it is the peak-to-peak noise from the blank injection which is used to calculate signal-to-noise for all peaks in all injections.

It’s that easy!

Final Notes

This can be done with either the Pro or QuickStart interface.