In this tip, we will learn ways to make working with the Method Validation Manager easier, particularly around Sample Set Methods and Sample Sets.
A Sample Set Method needs to be attached to each test when creating the validation protocol. Likewise, a Sample Set must be attached to each test when conducting a validation study.
Validation Protocols
A Sample Set Method from a non-validation project can be copied into a validation template project.
Once the Sample Set Method has been copied into the validation template project, open the method, and select which injections will be used for which tests. Once that is done you can attach the Sample Set Method to the appropriate test(s).
You can create a Sample Set Method within the Validation Manager Window by selecting File, New, Sample Set Method. Once the method has been created you will need to select which injections will be used for which tests and the new Sample Set Method can be attached to the appropriate test(s).
Validation Studies
Likewise, a Sample Set which was collected in a non-validation project can be copied into a validation working project. You will need to alter the Sample Set to indicate which injections are to be used for which tests before attaching the Sample Set to the appropriate test(s).
A Result Set can be copied from a non-validation project into a validation working project. This too would need to be altered to indicate which injections go with which tests. The Result Set would need to be reprocessed and you could select Use Existing Integration. Once that has been done you can attach the Result Set to the appropriate test(s).
Using the Pro interface when working with the Method Validation Manager option is recommended.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
Serve Aiuto? Contattate un esperto Waters.