In this tip for Empower Chromatography Data System (CDS), we will answer an Empower user question.
QUESTION:
How does Empower calculate the %Deviation for points in the calibration curve table?
The answer depends on how a particular System Policy is set.
There is a System Policy titled “Calculate % Deviation of Point From Curve”, found on the Data Processing tab in System Policies. By default, that System Policy is NOT enabled.
STEP 2:
When the policy is NOT enabled the % Deviation is calculated as seen in figure 2.
If we look at the first point on the calibration curve, this results in a % Deviation of -6.542.
What happens when the System Policy IS enabled?
When the policy IS enabled the % Deviation is calculated as seen in figure 5.
In this case, the first point on the calibration cures results in a % Deviation of 7.000.
It’s that easy!
Both Tangential and Exponential Skims can skim a rider either from the front of the main peak or the back of the main peak.
This can be done with either the Pro or QuickStart interface.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
支援をご希望ですか?ウォーターズの専門家に相談しましょう。