Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Tip 413

Tip 413

Empower Tip: ALCOA++

Empower Tip: ALCOA++

Tip #413: Configuring Empower for ALCOA++ | Legible

In this tip, we learn how to configure Empower for Legible.

Legible means that the data is legible and permanent; it includes use of secure, time stamped audit trails. (Note: I recommend reading Tip 411 before reading this tip)

Let’s see how it’s done!

Things to consider:

Set New Project Audit Trail System Policies:

Full Audit Trail Support enabled

Changes to GXP data need a reason

Silent – no comment required

Unrestricted – enter free text

Restricted – select from a list of reasons which are Default Strings

Confirm Identity forces reauthentication before the action occurs

As set, ALL projects will be audit trailed with no option to create a new project without audit trail:

For the System Audit Trail:

Items not used or not directly impacting GxP do not have to have a reason

Can be useful for system management to record reasons around Systems, Nodes, Users, User Types and so forth

Highest risk items may need reauthentication

It’s that easy!

Note: these settings can be made in System Policies in Configuration Manager.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



支援をご希望ですか?ウォーターズの専門家に相談しましょう。 

トップに戻る トップに戻る