Question: How do you summarize specified and unspecified impurities in a sample and quantitate them against the main component?
Another great question.
We begin by opening a Processing Method and selecting the Components tab. In the Component Type field, select either specified impurity or unspecified impurity for the peaks of interest.
In the Curve Reference field, select the main component for the impurities so that Empower quantitates the impurities against the calibration curve for the main component. The Relative Response field is used to multiply the calculated amount by a factor, the default is 1.
On the Impurity tab, set the Impurity Response to Amount and select the Main Component. Then create two User Defined Impurity Groups – one for unspecified impurities and one for specified impurities. Enter the names for the groups and select the Component type from the drop down list.
Quantitate the sample and we see the total Impurity Response for the specified impurities and the total Impurity Response for the unspecified impurities.
Scroll to the right in the Peaks table to see the thresholds set in the Processing Method.
It’s that easy!
This procedure can be followed using the QuickStart or Pro interface.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
支援をご希望ですか?ウォーターズの専門家に相談しましょう。