Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower TIp: Review Window and the Processing Method

Empower TIp: Review Window and the Processing Method

Tip #109: How to track peaks using the ACQUITY QDa mass detector

Tip #109: How to track peaks using the ACQUITY QDa mass detector

In this tip we are going to learn about peak tracking. A typical reversed-phase method development scenario would include trying different columns and varying the strong solvent for the mobile phase. The challenge is tracking the peaks from experiment to experiment. We could build a library of UV spectra for our compounds of interest with a PDA and then have Empower match the apex spectrum of each peak to the spectra in the library. That works well if the UV spectra are sufficiently different for the compounds of interest. An alternate approach would be to track the peaks by m/z ratio using the ACQUITY QDa mass detector. We can then summarize peak tracking using Assigned Mass.

STEP 1 On the MS 3D Channel tab of the Processing Method, peak tracking is enabled by first selecting ‘Apex’ or ‘Combined’ from the Spectra for Assigned Mass drop-down list. (Refer to Tip 103 for an explanation of Combined)

STEP 2

STEP 2

Set the desired precision.

STEP 3

STEP 3

Save the changes to the Processing Method and process the Sample Set. Create a ‘Summary By All Report Method’ that includes the Peak Track Component Summary table.

STEP 4

STEP 4

This summarizes the peaks by Assigned Mass. We can track the peaks as we vary the chromatographic conditions. For example, we see mass ‘182’ and the retention time for the peak with that mass over eight experiments.

STEP 5

STEP 5

Now let’s look at the Result Set. In the first result, we see the ‘Assigned Mass’, ‘Assigned Mass Value’, and the ‘Notes’ fields have been added to the Peaks table. The Base Peak value from either the Apex Spectrum or the Combined Spectrum is then copied into the ‘Assigned Mass Value’ field using the precision set in the Processing Method.

STEP 6

STEP 6

In this result, we see that peaks 6 & 7 are fused and have the same Assigned Mass. Empower has appended these Assigned Masses with numbers. Now let’s evaluate the assignment.

STEP 7

STEP 7

In the Mass Analysis window, we look at the Purity View for these two peaks. (Refer to Tip 103 for an explanation of Purity view). We may decide that the peak at 2.483 minutes has a mass of 196 and that the 202 comes from the peak at 2.465 minutes.

STEP 8

STEP 8

Back in the main window, ‘right-click’ on the row for the peak 2.483 minutes and select ‘Modify Assigned Mass’.

STEP 9

STEP 9

Enter the mass and any notes, then click OK.

STEP 10

STEP 10

The peak table has been updated. Save the result.

STEP 11

STEP 11

Summarizing the Result Set once again, we see the new summary which includes the manual assignment.

It’s that easy!

Final Notes

For more detailed information, read the Application Note: Streamline Method Development with Empower 3 MS Peak Tracking.

This procedure can be followed using the QuickStart or Pro interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



도움이 필요하십니까? Waters 전문가에게 문의하기

맨 위로 돌아가기 맨 위로 돌아가기