Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Peak Purity

Empower Tip: Peak Purity

Tip #251: Considerations for Collecting Data Suitable for Peak Purity Determination

Tip #251: Considerations for Collecting Data Suitable for Peak Purity Determination

In this tip, we begin a new series on Peak Purity using both the ACQUITY™ PDA Detector and ACQUITY QDa™ Mass detector. We will discuss Peak Purity and the considerations for collecting data suitable for Peak Purity determination?

STEP 1

STEP 1

Peak Purity is spectral purity or spectral homogeneity. Peak Purity is not chemical purity. If we extract spectra across a peak and normalize them, do they overlay? If YES, chances are there is one component under the peak. If NO, there are multiple components under the peak. We can use UV spectra from the ACQUITY PDA Detector

or MS spectra from the ACQUITY QDa Mass Detector to evaluate Peak Purity.

STEP 2

STEP 2

The following parameters in the ACQUITY PDA Instrument Method are critical when evaluating Peak Purity.

Wavelength Range: Start above the UV cutoff of the mobile phase and end at a wavelength that encompasses all areas of absorbance. If the mobile phase has a solvent or buffer that absorbs at the low end, the absorbance by the mobile phase will lower the linear dynamic range of the detector by the amount of absorbance that is autozeroed out

Resolution: Select 1.2nm which will give you the best possible spectral resolution

Sampling Rate: Select a rate that will provide at least 12 spectra across the narrowest peak of interest, 15-20 spectra is optimum.

STEP 3

STEP 3

The following parameters in the ACQUITY QDa Mass Detector Instrument Method are critical when evaluating Peak Purity

Mass Range: Select an appropriate range

Sampling Rate: Select a rate that will provide 15-20 spectra across the narrowest peak of interest. Mass Detectors do adjust the Sampling Rate in the instrument method based on the mass range that you enter. That means for large mass ranges the software automatically adjusts the Sampling Rate to a lower value – and the reverse is true, i.e. for a smaller range it will calculate a higher Sampling Rate.

STEP 4

STEP 4

Keep the absorbance of all peaks less than 1 AU according to MaxPlot. Why? At high absorbances a combination of effects can produce slight departures (approx. 1%) from Beer’s law because of photometric error. Although these errors at this level have a negligible effect on quantitation, they can be a source of spectral inhomogeneity and result in distortion of spectra.

Therefore, to minimize the effects of photometric error for the spectral contrast math used to determine Peak Purity, the maximum spectral absorbance of a compound should be less than 1.0AU.

STEP 5

STEP 5

A 5% error in linearity can have a profound effect on Peak Purity results.

It’s that easy!

Final Notes

This can be done with either the Pro or QuickStart interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



도움이 필요하십니까? Waters 전문가에게 문의하기

맨 위로 돌아가기 맨 위로 돌아가기