Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Calculating Amount/Concentration

Empower Tip: Calculating Amount/Concentration

Tip #267: How Empower calculates amount/concentration

Tip #267: How Empower calculates amount/concentration

In this tip for Empower Chromatography Data System (CDS), we begin a series on how Empower calculates Amount/Concentration and discuss the sections of the Sample Set Method and the Processing Method which are needed for the calculation of Amount/Concentration.

Let’s begin this series with a simple scenario. In this example, we want to do external standard calibration with multiple standards each with different amounts of a compound and then quantitate our samples against the calibration curve.

STEP 1

STEP 1

The Sample Set has five standards and three samples. In this example, we leave the Sample Weight and Dilution fields at ‘1’ which are the default values. With those fields set to ‘1’ they have no effect on the calculations and we will discuss those fields in the next tip.

STEP 2

STEP 2

The Component Editor displays ‘1’ component, ‘Caffeine’, with increasing amounts from standard ‘1 thru standard 5’. We are not using the Purity tab and will discuss that in the next tip.

STEP 3

STEP 3

The Processing Method displays one line for the component ‘Caffeine’. The Sample Value Type is ‘Amount’. For the calibration curve the x axis is ‘Amount’, the y axis is ‘Area’, and the fit type is ‘Linear’.

STEP 4

STEP 4

The Sample Set was Processed with the Processing Method. The Results window within Review displays the information about the calibration in the lower table. The equation for the line along with the A & B coefficients is displayed.

STEP 5

STEP 5

The Calibration Curve window within Review displays the calibration curve along with the point table. Because we selected a ‘Linear’ fit in the Processing Method, Empower performs a linear fit of the points.

STEP 6

STEP 6

The calculated value in the point table can be determined by rearranging the equation for a line and solving for x

STEP 7

STEP 7

The ‘% Deviation’ calculation depends on the selection of the highlighted System Policy. By default, this System Policy is not enabled and therefore, Empower uses ‘equation 1’. If you enable the System Policy, Empower uses ‘equation 2’.

STEP 8

STEP 8

The sample is quantitated, and the amount of Caffeine is calculated and displayed in the Peaks table. Once again, the amount is calculated solving for ‘x’ in slide 6.

It’s that easy!

Final Notes

This can be done with either the Pro or QuickStart interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



도움이 필요하십니까? Waters 전문가에게 문의하기

맨 위로 돌아가기 맨 위로 돌아가기