Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Calculating Amount/Concentration

Empower Tip: Calculating Amount/Concentration

Tip #283: Calibration Curve Fit Types: Fifth Order Fit

Tip #283: Calibration Curve Fit Types: Fifth Order Fit

In this tip for Empower Chromatography Data System (CDS), we continue our series on Calculating Amount/Concentration, Calibration Curve fit types, and review Firth Order Fit.

STEP 1

STEP 1

For a Fifth Order Fit, Empower performs a least-squares fit of a fifth-order polynomial to the calibration points. This fit cannot be performed with less than 6 points and 8 is recommended.

STEP 2

STEP 2

The Fifth Order Fit generates a Calibration Curve that is represented by the equation in figure 2. In this example, it is a GPC Processing Method that was used for a 15-point Calibration Curve.

STEP 3

STEP 3

Calibration Curves can be accessed in various ways, for example:

Review – bring a result into Review and click on the Calibration Curve tool

Compare – right click on a curve from the Curves tab in the project and select Compare

It’s that easy!

Final Notes

This can be done with either the Pro or QuickStart interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



도움이 필요하십니까? Waters 전문가에게 문의하기

맨 위로 돌아가기 맨 위로 돌아가기