In this tip for Empower Chromatography Data System (CDS), we continue our series on the upcoming changes to USP Chapter 621 for Chromatography.
In this tip, we will discuss how to comply with USP requirements using your current version of Empower.
To recap the changes from Tip 301, they are as follows:
Resolution, Relative Resolution and Plate Count will use width at Half Height
Width at Tangent is no longer used for any calculation
Plate Count will be called Plate Number
Tailing Factor will be called Symmetry Factor, no change to the calculation
The Half Height Multiplier for signal-to-noise will change from 5 to 20, no change to the calculation
There are fields already calculated in Empower which you can report to comply with the changes using the version of Empower you have today. You can rename them accordingly:
USP Resolution (HH), Resolution per both the EP and JP all use peak width at half height
Relative Resolution uses peak width at half height
EP Plate Count and JP Plate Count use peak width at half height
USP Tailing, Symmetry Factor per both the EP and JP
An alternative for the calculation of Resolution is to create a Custom Field. Please remember that any Custom Field should be validated before putting it into routine use.
An alternative for the calculation of Plate Count is to create a Custom Field. Again, please validate the Custom Field before it is put into routine use.
An example is shown reporting USP Resolution (HH), EP Plate Count and USP s/n.
The Half Height Multiplier has been changed from 5 to 20 in the Processing Method to comply with the new requirement.
It’s that easy!
This can be done with either the Pro or QuickStart interface.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
도움이 필요하십니까? Waters 전문가에게 문의하기