Empower provides two ways to determine if a specified mass exists in a sample. You may be looking for a specific mass because it identifies a desired ‘component’ or an ‘impurity’ in the sample. Today, we will look at ‘component-based’ Expected Mass Processing.
STEP 1 Bring the Sample Set (or injections) into Review and open and apply the Method Set. Go to the Processing Method and click the ‘MS Expected Mas’ tab.
Select ‘Component-based’ processing and enter the Expected Mass or Masses along with the Expected Intensities. In this case, the compound of interest has a molecular ion with a base peak of 621, a fragment ion with mass of 603, a sodium adduct ion with mass of 643, and isotopic ions with masses of 622 and 644. The Peak Separation has a default value of 1 and sets the range over which Empower looks for the expected mass. For example, for mass 621 the range would be 620.5 to 621.5. The expected intensity is a % of the base peak and if the % is equal to or higher than the entered value it is acceptable.
In this example, all expected masses are ‘found’ and the expected intensities are ‘acceptable’.
In this example, some of the expected masses are ‘not found’ and the expected intensities are ‘below the value’ entered in the Processing Method, therefore, Empower flags the peak.
It’s that easy!
This procedure can be followed using the QuickStart or Pro interface.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
Precisa de ajuda? Conecte-se a um especialista da Waters.