In this tip for Empower Chromatography Data System (CDS), we will learn about calculations for Gradient Polymer Elution Chromatography (GPEC).
Let’s start with some background information. GPEC is an LC technique. In this example, chromatograms in Figure 1, three chemically different polymers co-elute using a GPC separation because they have similar hydrodynamic volumes. If reverse phase separation is used to separate these polymers, they would remain partially soluble throughout the separation and would not be baseline separated. The GPEC method purposefully precipitates the polymers onto the column or column frit until the gradient ratio of good solvent reaches the polymers solubility point, thereby releasing the polymer from the surface of the column. Instead of using the size-based separation of GPC, the polymers are separated by their chemical solubility using GPEC.
Practical examples where GPEC is helpful:
Reverse engineering
Batch to batch conformity
Optimization of co-extrusion or formulation
Aligning thermal properties with a chemical analysis
Our goal is to calculate the percentages of solvent used in the mobile phase to separate the polymers.
We begin with the gradient table in the Instrument Method. The method starts with 100% methanol and ends at 100% THF.
On the Data tab of the Instrument Method, select the solvent lines with the solvents used in the gradient.
The polymers are separated, and we record the retention time of each.
Peak Custom Fields are created to report the ‘y’ value at the retention times determined in Step 3.
The ‘%A’ channel has been processed and we see the ‘%methanol’ at which each polymer elutes.
The ‘%C’ channel has been processed as well and we the ‘%THF’ at which each polymer elutes.
It’s that easy!
This can be done with either the Pro or QuickStart interface.