Tip #314: Creating a Custom Field to calculate the %Area of multiple groups of peaks in a chromatogram

Tip #314: Creating a Custom Field to calculate the %Area of multiple groups of peaks in a chromatogram

In this tip for Empower Chromatography Data System (CDS), we will answer an Empower user question.

QUESTION:

I have multiple groups of peaks in the chromatogram. Each group has known peaks at the start and end of each group. I want to calculate the %Area of each group not including the known peaks which border each group. Furthermore, the peaks shift in retention time from run to run. How can I accomplish this?

Great question. The short answer is to create a Custom Field which adds up the areas of the peaks in between the known peaks, divides it by the total area (automatically calculated by Empower), and multiplies by 100.

STEP 1

STEP 1

In this example, there are two groups, the first between Peaks A and B, and the second between Peaks C and D.

STEP 2:

Make sure the known peaks are identified properly, therefore, use the RT Window, Peak Match, and if necessary, the RT Reference and the RRT fields. (Note: Refer to Tip 310 for more details on using RT Reference and RRT)

STEP 3

STEP 3

Create one Result Custom Field per group.

Group 1 formula:

SUM(GT(Retention Time,A[Retention Time])&LT(Retention Time,B[Retention Time])*Area)/Total Area*100 .

STEP 4

STEP 4

Group 2 formula:

SUM(GT(Retention Time,C[Retention Time])&LT(Retention Time,D[Retention Time])*Area)/Total Area*100

STEP 5

STEP 5

The results can be viewed in the Results window within Review.

It’s that easy!

Final Notes

This can be done with either the Pro or QuickStart interface.