In this tip, we will answer an Empower user question.
QUESTION:
Is it normal for my blank to be flagged when a Must Peak is absent?
Great question. Yes, it is normal behavior if a blank is flagged due to a missing Must Peak, Default Peak, RT Reference Peak, or Internal Standard Peak.
Let’s review.
This Sample Set has a blank, injected 3 times.
On the Component tab of the Processing Method, Peak A is selected as a Must Peak. This means that Peak A must be found in every chromatogram.
Peak A is not found in the blank, therefore, the result is flagged as Faulted and Result Code F01 is displayed.
A similar situation is observed if an Internal Standard peak, a Default peak, and/or an RT Reference peak is missing. Like a Must Peak, these types of peaks must also be present in every chromatogram. The online Help defines the Result Codes that indicate if one of these peaks is missing.
It’s that easy!
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
Precisa de ajuda? Conecte-se a um especialista da Waters.