Tip #413: Configuring Empower for ALCOA++ | Legible
In this tip, we learn how to configure Empower for Legible.
Legible means that the data is legible and permanent; it includes use of secure, time stamped audit trails. (Note: I recommend reading Tip 411 before reading this tip)
Let’s see how it’s done!
Things to consider:
Set New Project Audit Trail System Policies:
Full Audit Trail Support enabled
Changes to GXP data need a reason
Silent – no comment required
Unrestricted – enter free text
Restricted – select from a list of reasons which are Default Strings
Confirm Identity forces reauthentication before the action occurs
As set, ALL projects will be audit trailed with no option to create a new project without audit trail:
For the System Audit Trail:
Items not used or not directly impacting GxP do not have to have a reason
Can be useful for system management to record reasons around Systems, Nodes, Users, User Types and so forth
Highest risk items may need reauthentication
It’s that easy!
Note: these settings can be made in System Policies in Configuration Manager.
Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.
Precisa de ajuda? Conecte-se a um especialista da Waters.