Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Question from an Empower User

Empower Tip: Question from an Empower User

Tip #165: Question: Does Empower provide warnings when an instrument is out of calibration?

Tip #165: Question: Does Empower provide warnings when an instrument is out of calibration?

In this tip, I am going to answer a great question I received from an Empower user.

QUESTION:

Does Empower provide warnings when an instrument is out of calibration?

Great Question. It is critical in a regulated environment that when the service date for an instrument is reached, users are prevented from using that instrument. Empower can do that.

STEP 1

STEP 1

In Configuration Manager, right-click on the Node to which the instrument is connected and select ‘Properties’.

STEP 2 Click the ‘Instruments’ tab. This screen displays what instruments are connected to the Node and their status.

STEP 3

STEP 3

Scroll to the right and locate the ‘Last Serviced’ field. This is where you (or your Waters Service Engineer) can enter the date when the instrument was last serviced.

STEP 4

STEP 4

Scroll further to the right and locate the ‘Next Service Due’ date. This is where you (or your Waters Service Engineer) can enter the next scheduled service date.

STEP 5

STEP 5

Go to ‘User Types’ and right click on the ‘User Type Name’ of the user and select ‘Properties’.

STEP 6

STEP 6

On the ‘Data Acquisition’ tab, locate the ‘Warn On Service Limit’ privilege. If the user has been given this privilege, then the user can use the instrument beyond the service date entered in Step 4. If the user does not have this privilege, then the user is prevented from using the instrument once that date is reached.

STEP 7

STEP 7

When the user without the privilege attempts to access ‘Run Samples’ with this instrument, a warning message appears. Once the user clicks ‘OK’, the Run Samples window will close.

STEP 8

STEP 8

The message center will display an error message for the user who tried to access that instrument.

It’s that easy!

Final Notes

This procedure can be followed using the QuickStart or Pro interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



您需要帮助吗?联系沃特世专家。 

返回顶部 返回顶部