Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Method Validation Manager

Empower Tip: Method Validation Manager

Tip #191: Validation Protocols: Setting up the Accuracy Test

Tip #191: Validation Protocols: Setting up the Accuracy Test

In this tip, we will continue our discussion on Validation Protocols and learn how to set up the Accuracy Test. (NOTE: As mentioned previously, you should follow your own SOP, this is only an example.)

STEP 1

STEP 1

Select the ‘Accuracy’ test from the ‘Test Name’ field in the upper table. In the ‘Test Status’ field, it indicates the test is ‘incomplete’ because it has not yet been configured. Highlight the row number for the test and click the ‘Test Parameters’ tool.

STEP 2

STEP 2

On the Acquisition tab the number of ‘levels’ ‘preparations per level’, and ‘injections per preparation’ are entered. In this example, the total number of injections is 9. From the ‘Levels’ field, Levels 1, 4 and 5 have been selected.

STEP 3

STEP 3

On the Processing and Acquisition tab, the ‘average for percent recovery’ has a target value of 100 and a range of 98 to 102. If the calculated value is ‘outside’ of the set range, it will be flagged and the test will fail. Additionally, we set an upper limit for the %RSD for the percent recovery at 2.000000. If the calculated value ‘exceeds’ the upper limit, it will be flagged and the test will fail.

STEP 4

STEP 4

Attach the appropriate Sample Set Method by selecting it from the Sample Set method field in the lower table. Empower checks that the number of injections in the Sample Set Method to match the number of injections specified in the test. There is agreement, hence the green check mark in the SSM Status field. Now that we have configured the test, in the upper table the ‘Test Status’ field indicates that the ‘Sample Sets are Incomplete’ because we have not yet acquired the data.

One Sample Set Method can be used for multiple tests. In the lower table, highlight the row number for the Sample Set Method and click the ‘Open/Edit Sample Set Method’ tool to view the Sample Set Method.

STEP 5

STEP 5

Note that the method contains the necessary injections for the three configured tests: Linearity, Repeatability, and Accuracy.

It’s that easy!

Final Notes

Please use the Pro interface when working with the Method Validation Manager option.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



您需要帮助吗?联系沃特世专家。 

返回顶部 返回顶部