Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Peak Purity

Empower Tip: Peak Purity

Tip #256: How to Calculate the Solvent Angle

Tip #256: How to Calculate the Solvent Angle

In this tip for Empower Chromatography Data System (CDS), we will discuss how to calculate the Solvent Angle.

Continuing the discussion regarding AutoThreshold from last week (Tip 255), if the Purity Angle is greater than the Purity Threshold for any of the peaks in any of the standard injections, then you need to determine the Solvent Angle separately.

STEP 1

STEP 1

On the Purity tab of the Processing Method, select ‘Noise’ from the ‘Threshold Criteria’ list.

STEP 2

STEP 2

Process the same six injections of the standard that was used to verify AutoThreshold.

[figures 2]

STEP 3

STEP 3

Select the ‘highest Purity Angle’ reported from the results.

STEP 4

STEP 4

Select ‘Noise+Solvent’ from the ‘Threshold Criteria’ list and enter the value from Step 3 into the ‘Solvent Angle (degrees)’ field. This value has a precision of 2.

STEP 5

STEP 5

Save all the changes to the Processing Method and reprocess the six injections. Now the Purity Angle should be less than the Purity Threshold for all peaks in the standards.

It’s that easy!

Final Notes

This can be done with either the Pro or QuickStart interface.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



您需要帮助吗?联系沃特世专家。 

返回顶部 返回顶部