Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: USP Chapter 621 for Chromatography

Empower Tip: USP Chapter 621 for Chromatography

Tip #304: A Few Common Questions Answered

Tip #304: A Few Common Questions Answered

In this tip for Empower Chromatography Data System (CDS), we conclude our series on the upcoming changes to USP Chapter 621 for Chromatography.

In this tip, we will conclude our series by answering a few commonly asked questions surrounding these changes.

QUESTION 1

With regards to the half height multiplier changing from 5 to 20, what if I do not have enough baseline or there are interfering peaks?

A: As per USP Chapter 621, if a baseline of 20 times the width at half-height is not obtainable because of peaks due to the solvents or reagents or arising from the mobile phase or the sample matrix, or due to the gas chromatographic temperature program, a baseline of at least 5 times the width at half-height is permitted.

QUESTION 2

What is the impact on equipment qualification, especially regarding the signal-to-noise (s/n) change as a requirement of the qualification process?

A: There are a variety of approaches to measuring signal-to-noise, especially the noise portion of the equation. For chromatographic system qualification, Waters does not believe that the upcoming USP changes in Chapter 621 have an impact. The USP definition of noise is intended specifically for USP monograph procedures and does not dictate the approach for other types of procedures, nor for chromatographic system qualification. The s/n specification defined for a given system and used in the Waters SystemsQT for Systems is based on factory specifications and testing using the defined s/n measurements and chromatographic conditions. This data is used to define the acceptance criteria for our SystemsQT for Systems protocols. Changing the approach to measuring s/n and redefining the specification and acceptance criteria may result in slightly different numbers; however, this does not in any way negate the validity of the current s/n calculation nor does it affect the performance of the system.

QUESTION 3

Is Waters developing a hotfix for older versions of Empower or any type of tool to automatically update Empower to comply with these changes?

A: There are no plans to provide hotfixes for these changes as these types of changes are not conducive to a simple turnkey solution in Empower. It may be possible for your local Professional Services teams to assist in evaluating the impact of these changes to your organization and to assist with necessary updates to Empower methods or other types of configuration changes. Please contact your local Waters representative if you would like to pursue this.

QUESTION 4

Has the date for the change of USP Chapter 621 been delayed?

A: The changes go into effect on 1 December 2022. What has been delayed are the subsections for System Sensitivity and Peak Symmetry.

QUESTION 5

Will I have to revalidate my methods once these changes take place and will this affect acceptance criteria?

A: I would think acceptance criteria would remain the same, i.e., baseline resolution between adjacent peaks is classically defined at >= 1.5. I do not think the methods will need revalidation; however, I strongly suggest you contact the USP with this question.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



您需要帮助吗?联系沃特世专家。 

返回顶部 返回顶部