Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: QDa MS Data in Empower

Empower Tip: QDa MS Data in Empower

Tip #357: Obtaining Base Peak Values from the UV Chromatogram

Tip #357: Obtaining Base Peak Values from the UV Chromatogram

In this tip, we will expand on Tip 356, where we now have an example where there is a PDA and QDa connected in series. Since we are comfortable working with PDA data, we can use the UV chromatogram to get base peak values from the peaks.

Let’s see how it’s done!

STEP 1

STEP 1

Bring the 3D PDA channel into Review and extract a wavelength of interest.

STEP 2

STEP 2

Integrate the peaks either manually or by using a processing method and the base peak field is populated for any integrated peaks and the UV spectra for those peaks appear in the Spectrum Review window.

STEP 3

STEP 3

Click the Mass Analysis tool to open the Mass Analysis window or select Mass Analysis Window from the Window menu. The UV and MS spectra are displayed for every integrated peak. The Mass Analysis window displays a variety of information which we will explore in future tips.

STEP 4

STEP 4

From the 2D Channels tab, we can select the MS TIC and overlay it with the UV chromatogram. (Be sure to click the Autoscale tool to correct for the difference in scaling.) You can see that the peaks in the MS chromatogram elute slightly later than those in the UV chromatogram because the QDa detector is second in the series.

STEP 5

STEP 5

Click on the MS chromatogram and you can see the MS spectra for the peaks in Spectrum Review. The retention times in the Spectra table are taken from the peaks in the UV chromatogram.

STEP 6

STEP 6

We can correct this slight offset in the Smoothing/Offset tab of the processing method by entering a small offset. It is recommended to calculate the average retention time difference for the peaks between the two channels and use that as the offset value.

STEP 7

STEP 7

Reprocessing the injection with the revised processing method gives us a better overlay of the chromatograms and the retention times of the spectra are corrected to correspond to the peaks in the MS chromatogram.

It’s that easy!


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



您需要帮助吗?联系沃特世专家。 

返回顶部 返回顶部