Moving the pump plunger backward

Moving the pump plunger backward
Topic LCI-USG-0195, 最近更新 05/01/2025

Use the touchscreen UI to move the pump plunger backward before replacing pump components.

To move the pump plunger backward:

  1. Flush the pump with nonhazardous solvent.
  2. From the touchscreen, select the solvent manager, and then tap Maintain > Replace Components.
  3. From the Replace Components page, scroll down and then tap Pump Maintenance.
  4. Tap Pump Head Maintenance.
  5. Select the plunger head you want to move backward.
  6. Tap Move Backward, and then wait for the plunger to stop.