Tip #107: Smoothing the chromatogram to improve the baseline and the signal-to-noise ratio

Tip #107: Smoothing the chromatogram to improve the baseline and the signal-to-noise ratio

In this tip we are going to learn why smoothing the chromatogram can be helpful (Part 7). If the baseline is noisy you can use a small amount of smoothing to improve the baseline and the signal-to-noise ratio. Please note: Be careful, smoothing algorithms smooths all data points, not just baseline points. If you apply too high of a smoothing factor you can smooth away your peaks.

STEP 1 We start without any smoothing and see the peak of interest has a signal-to-noise ratio of 29:1.

STEP 2

STEP 2

There is more information if you go to the Results window within Review. We see the height, signal-to-noise ratio, and the peak-to-peak noise which is 33659.

STEP 3

STEP 3

Now we apply a small amount of smoothing using the Smoothing/Offset tab in the Processing Method.

STEP 4

STEP 4

The signal-to-noise ratio for our peak of interest is now 42:1.

STEP 5

STEP 5

Returning to the Results window, we see a slight reduction in peak height and a significant reduction in the peak-to-peak noise to 22715.