Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Empower User Question

Empower Tip: Empower User Question

Tip #210: Reporting Custom Field Values in Empower

Tip #210: Reporting Custom Field Values in Empower

In this tip, we will answer another Empower user question regarding Custom Field values.

QUESTIOM:

I have an Intersample Peak Custom Field with a summary function which works as expected. The result of that Custom Field is then used in a second Peak Custom Field which works as expected. The result of the second Custom Field is then used in a third Intersample Peak Custom Field with a summary function. The result of that Custom Field increments until the last injection. At the last injection the proper value is reported. How can I get Empower to report the final value without reporting incremental values?

STEP 1

STEP 1

Custom Field 1 is a simple Intersample Peak Field calculating the average area for the peaks using all injections.

STEP 2

STEP 2

Custom Field 2 is a Simple Peak Field that takes the average area, minus the area for an individual peak, divided by the average area, times 100.

STEP 3

STEP 3

Custom Field 3 is a simple Intersample Peak Field calculating the average value for Custom Field 2.

STEP 4

STEP 4

Custom Fields 1 and 2 are calculated properly. Custom Field 3 increments, meaning the average for injection 1 of peak 1 is simply the value of Custom Field 2. The average for injection 2 of peak 1 is the average value of Custom Field 2 for the first 2 injections. Finally, the average for injection 3 of peak 1 is the true average value for all 3 injections.

STEP 5

STEP 5

The solution is to include the full calculation for Custom Field 2 in Custom Field 3.

STEP 6

STEP 6

Now the average value reported is the same for all injections and the proper values are reported.

It’s that easy!

Final Notes

You can use either the Pro or QuickStart interface for this tip.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



需要協助嗎?聯絡Waters專家。 

回到頁首 回到頁首