Do you want Empower Tips sent to your inbox each week?
  • Support Articles

Empower Tip: Review Window and the Processing Method

Empower Tip: Review Window and the Processing Method

Tip #31: Calculating signal-to-noise ratio (Part 4).

Tip #31: Calculating signal-to-noise ratio (Part 4).

STEP 1

STEP 1

Let’s start by creating the custom calculation. This is done in the properties of your project. Go to the Configuration Manager window by clicking Configure the System from the Pro Interface.

STEP 2

STEP 2

Right-click on the row number for your project and select Properties.

STEP 3

STEP 3

Click the Custom Fields tab and then click the New button to launch the New Custom Field Wizard.

STEP 4

STEP 4

On the Data and Type Selection screen select Peak for the Field Type and Real for the data type. A peak custom field is a calculation that can be applied to every peak in the chromatogram. Click Next.

STEP 5

STEP 5

Do not change anything on the Source Selection screen. Click Next.

STEP 6

STEP 6

We will enter the formula on the Formula Entry screen. Signal-to-noise ratio is peak height divided by detector noise. In the Fields list scroll to the right and double-click on Height.

STEP 7

STEP 7

From the Operations list double click the forward slash which is the division symbol.

STEP 8

STEP 8

Height is in microvolts and detector noise is in plot units so we will need to correct for the difference in units. In the denominator enter a left parenthesis.

STEP 9

STEP 9

In the Fields list locate Detector Noise and double click to add it to the formula. Then double-click the forward-slash.

STEP 10

STEP 10

In the Fields list locate Scale to uV and double-click to add it to the formula. The Scale to uV will correct for the difference in units. Then add the right parenthesis. Click Next.

STEP 11

STEP 11

Set the desired precision on the Numeric Parameters screen and click Next.

STEP 12

STEP 12

Enter a name on the Name Entry screen and click Finish. Note: you cannot use spaces in the name but could do shift underscore.

STEP 13

STEP 13

Click OK on the warning message.

STEP 14

STEP 14

Now let’s use the custom field. Open the project and bring raw data (Sample Set, Injection, or Channel) into review and open the Processing Method. Click the Processing Method tool.

STEP 15

STEP 15

Click the Noise and Drift tab.

STEP 16

STEP 16

Check the box for Calculate Detector Noise and Drift and enter a time interval which is free of peaks for calculating noise. This time region must be equal to or greater than the Segment Width and must contain at least 60 data points. The Segment Width is used for calculating average noise and drift values and it must contain at least 30 data points.

STEP 17

STEP 17

Return to the Main Window and click the Integrate tool.

STEP 18

STEP 18

If signal-to-noise ratio is not visible we can add it to the Peaks table. Right-click on the Peaks table and select Table Properties.

STEP 19

STEP 19

On the Columns tab, find signal-to-noise ratio and remove the check mark to make it visible in the table. Click OK.

STEP 20

STEP 20

The signal-to-noise ratio is displayed for every integrated peak.

It’s that easy!

Final Notes

This procedure can be followed using the QuickStart interface. In step 1 described above, select Configuration from the Manage menu.


About the Author

About the Author
Neil Lander

Neil Lander

Neil provides internal support for Empower CDS software and is focused on developing laboratory software solutions that help organizations achieve their scientific and operational objectives.



需要協助嗎?聯絡Waters專家。 

回到頁首 回到頁首