Temperature

Temperature
Topic LCI-SPG-0010, Last Updated 07/15/2025

The ambient temperature in the laboratory must be from 4 to 40 °C (39 to 104 °F). Short-term thermal variations should be no more than 2 °C (3.6 °F) per hour.

Note: Failure to operate in this range will compromise system performance and can result in instrument failure.

The optimum temperature range is 19 to 22 °C (66 to 72 °F).