Solvent miscibility

Solvent miscibility
Topic LCI-USG-0046, 最近更新 05/01/2025

Before you change solvents, consult this table to determine solvent miscibility.

Be aware of the following effects and considerations:

  • Temperature affects solvent miscibility. If you are running a high-temperature application, consider the effect of the higher temperature on solvent solubility.
  • Buffers dissolved in water can precipitate when mixed with organic solvents.
  • You can make changes involving two miscible solvents directly. Changes involving two solvents that are not totally miscible (for example, from chloroform to water) require an intermediate solvent like n-propanol.
  • When you switch from a strong buffer to an organic solvent, thoroughly flush the system with distilled water before you add the organic solvent.
Note: λ cutoff is the wavelength at which the absorbance of the solvent equals 1 AU.
Table 1. Solvent miscibility
Solvent Polarity index Viscosity cP, 20 °C (at 1 atm) Boiling point °C (at 1 atm) Miscibility number (M) λ cutoff (nm)
ACN 6.2 0.37 81.6 11, 17 190
Acetic acid 6.2 1.26 117.9 14 n/a
Acetone 5.4 0.32 56.3 15, 17 330
Benzyl alcohol 5.5 5.80 205.5 13 n/a
DMF 6.4 0.90 153.0 12 n/a
DMSO 6.5 2.24 189.0 9 n/a
Ethanol 5.2 1.20 78.3 14 210
n-Hexane 0.0 0.313 68.7 29 n/a
MeOH 6.6 0.60 64.7 12 210
Methoxyethanol 5.7 1.72 124.6 13 n/a
1-Propanol 4.3 2.30 97.2 15 210
2-Propanol 4.3 2.35 117.7 15 n/a
THF 4.2 0.55 66.0 17 220
Triethylamine 1.8 0.38 89.5 26 n/a
Water 9.0 1.00 100.0 n/a n/a