Replacing the pressure transducer

Replacing the pressure transducer
Topic LCI-USG-0227, 最近更新 04/30/2025

Users or Waters field service personnel can replace the pressure transducer in the event of mechanical or electronic failure.

Warning: Observe Good Laboratory Practice (GLP) at all times, particularly when working with hazardous materials. Consult the Safety Data Sheets regarding the solvents you use. Additionally, consult the safety representative for your organization regarding its protocols for handling such materials.
Warning: To avoid personal contamination from contact with biologically hazardous or toxic materials, do not attempt to force or draw liquid or gas through the valve's inlet or outlet ports.
Note: To avoid contaminating system components, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Requirement: Wear clean, chemical-resistant, powder-free gloves when performing this procedure.

Required tools and materials

  • Chemical-resistant, powder-free gloves
  • 1/4-inch open-end wrench
  • 5/16-inch open-end wrench
  • T10 TORX driver
  • Replacement pressure transducer assembly

To replace the pressure transducer:

  1. Flush the pump with nonhazardous solvent that is miscible with the present mobile phase.
  2. Note: To avoid damaging electrical components and circuitry, do not disconnect an electrical assembly while electrical power is applied to a module. To completely interrupt power, set the on/off switch to the "off" position, and then disconnect the power cord from the ac source. Wait 10 seconds before disconnecting an assembly.
    Power-off the system. See Powering-off the system
  3. Open the pump’s door.
  4. Disconnect the pressure transducer cable (arrow) from the pressure transducer cable receptacle.
    Figure 1: Pressure transducer cable
  5. Use the 1/4-inch open-end wrench to disconnect the tubing fittings (arrows) at the top and bottom of the pressure transducer.
    Figure 2: Removing pressure transducer tubing connections
  6. Use the T10 TORX driver to remove the screws (arrows) and remove the pressure transducer.
    Figure 3: Pressure transducer screws
  7. Use the T10 TORX driver to tighten the screws (arrows).
  8. Reattach the tubing fittings (arrows) to the transducer, tighten them with your fingers to the extent possible, and then use the 1/4-inch open-end wrench to tighten the fittings an additional 1/6-turn (for existing fittings) or 3/4-turn (for new fittings).
    Figure 4: Reattaching pressure transducer tubing connections
  9. Reconnect the pressure transducer connector to the pressure transducer cable receptacle.
    Figure 5: Pressure transducer cable
  10. Power-on the system. See Powering-on the system.
  11. On the touchscreen, tap Health > Troubleshoot, tap the pump icon , and then tap System Leak Test.
    Tip: If the leak test results are unsatisfactory, pressurize the seals to properly seat them. To do so, run the pump at 58,605 kPa (586 bar, 8500 psi) for 30 minutes, or run the leak test until results are satisfactory.
  12. Follow the remaining on-screen prompts to run a system leak test.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

有關最新資訊,請參閱此頁面的英文版。

回到頁首 回到頁首