Replacing the PDA detector's detector lamp

Replacing the PDA detector's detector lamp
Topic LCI-USG-0215, 最近更新 05/01/2025

Replace the PDA detector's lamp annually during the prescribed preventive maintenance (PM) or any time it repeatedly fails to light or when the detector fails to calibrate. The system automatically senses the lamp upon installation, and its serial number and installation date are automatically recorded in the Lamp Change Record table.

Note: Waters warrants 2000 hours of lamp life, or one year from date of purchase, whichever comes first.
Warning: To prevent burn injuries, allow the lamp to cool for 30 minutes before removing it. The lamp housing gets extremely hot during operation.
Warning: To avoid eye injury from ultraviolet radiation exposure:
  • Power-off the detector before changing the lamp.
  • Wear eye protection that filters ultraviolet light.
  • Keep the lamp in the housing during operation.

To replace the lamp:

Caution: Ensure that you allow the lamp to cool for at least 30 minutes before starting this procedure.
  1. Power-off the system, and then disconnect the power cord.
  2. Open the PDA detector's door.
  3. Locate the lamp assembly cover (see the following figure).
  4. To remove the lamp assembly cover, use a Phillips screwdriver to remove the one Phillips-head screw (see the following figure).
  5. Lift the lamp assembly cover off.
    Figure 1: Lamp assembly cover
    Lamp assembly cover
    Phillips-head screw
  6. Disconnect the power cable to the lamp.
  7. Loosen the two captive screws in the lamp base.
    Figure 2: Lamp assembly base captive screws
    Captive screws
    Lamp cable
  8. Lift out the lamp assembly from the lamp housing.
    Note: To prevent shattering the glass, use care when disposing of the lamp because the lamp gas is under slight negative pressure.
    Figure 3: Removing the lamp assembly
  9. Position the new lamp so the cut-out located on the lamp base is at the 11 o’clock position, in line with the alignment pin on the lamp housing.
    Figure 4: Aligning the lamp
    Cut-out on lamp base at 11 o’clock position
    Alignment pin on lamp housing
  10. Gently push the lamp forward until it bottoms into position.
  11. Tighten the two captive screws.
  12. Reconnect the lamp’s power connector.
  13. Reinstall the lamp assembly cover using the one Phillips-head screw.
  14. When ready to resume operation of the PDA detector, reconnect the power cord, and then power-on the system.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

有關最新資訊,請參閱此頁面的英文版。

回到頁首 回到頁首