Consult this table for the properties of some common chromatographic solvents.
| Solvent | Vapor pressure mm Hg (Torr) | Boiling point (°C) | Flash point (°C) |
|---|---|---|---|
| ACN | 88.8 at 25 °C | 81.6 | 6 |
| Acetone | 184.5 at 20 °C | 56.29 | -20 |
| n-Butyl acetate | 7.8 at 20 °C | 126.11 | 22 |
| n-Butyl alcohol | 4.4 at 20 °C | 117.5 | 37 |
| n-Butyl chloride | 80.1 at 20 °C | 78.44 | -9 |
| Chlorobenzene | 8.8 at 20 °C | 131.69 | 28 |
| Chloroform | 158.4 at 20 °C | 61.15 | n/a |
| Cyclohexane | 77.5 at 20 °C | 80.72 | -20 |
| Cyclopentane | 400 at 20 °C | 49.26 | -7 |
| Dimethylacetamide (DMA) | 1.3 at 25 °C | 166.1 | 70 |
| DMF | 2.7 at 20 °C | 153.0 | 58 |
| DMSO | 0.6 at 25 °C | 189.0 | 88 |
| o-Dichlorobenzene | 1.2 at 20 °C | 180.48 | 66 |
| Dichloromethane | 350 at 20 °C | 39.75 | n/a |
| 1, 4-Dioxane | 29 at 20 °C | 101.32 | 12 |
| Ethyl acetate | 73 at 20 °C | 77.11 | -4 |
| Ethyl alcohol | 43.9 at 20 °C | 78.32 | 15 |
| Ethyl ether | 442 at 20 °C | 34.55 | -45 |
| Ethylene dichloride | 83.35 at 20 °C | 83.48 | 13 |
| Heptane | 35.5 at 20 °C | 98.43 | -4 |
| Hexane | 124 at 20 °C | 68.7 | -22 |
| IPA | 32.4 at 20 °C | 82.26 | 12 |
| Iso-octane | 41 at 20 °C | 99.24 | -12 |
| Isobutyl alcohol | 8.8 at 20 °C | 107.7 | 28 |
| Isopropyl myristate | <1 at 20 °C | 182.6 | 164 |
| MeOH | 97 at 20 °C | 64.7 | 11 |
| Methyl t-butyl ether | 240 at 20 °C | 55.2 | -28 |
| Methyl ethyl ketone | 74 at 20 °C | 79.64 | -9 |
| Methyl isobutyl ketone | 16 at 20 °C | 117.4 | 18 |
| N-Methylpyrrolidone | 0.33 at 25 °C | 202.0 | 86 |
| Pentane | 420 at 20 °C | 36.07 | -49 |
| n-Propyl alcohol | 15 at 20 °C | 97.2 | 23 |
| Propylene carbonate | n/a | 241.7 | 135 |
| Pyridine | 18 at 25 °C | 115.25 | 20 |
| TEA | 57 at 25 °C | 89.5 | -9 |
| TFA | 97.5 at 20 °C | 71.8 | -3 |
| THF | 142 at 20 °C | 66.0 | -14 |
| Toluene | 28.5 at 20 °C | 110.62 | 4 |
| 1,2,4-Trichlorobenzene | 1 at 20 °C | 213.5 | 106 |
| Water | 17.54 at 20 °C | 100.0 | n/a |
| o-Xylene | 6 at 20 °C | 144.41 | 17 |