Preparing and loading samples

Preparing and loading samples
Topic LCI-USG-0187, 最近更新 05/01/2025

The sample manager holds three ANSI/SBS-standard plates or trays that are loaded through the sample compartment door. Specific ANSI standard well plates, vial trays, vials, and cap mats or sealing caps are approved and required for use with the system. Incorrect loading of the plates or trays can result in an error.

Warning: To avoid personal contamination with biologically hazardous or toxic compounds, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid eye injury, use eye protection when performing this procedure.

Required tools and materials

  • Chemical-resistant, powder-free gloves
  • Protective eyewear

To prepare and load samples:

  1. Prepare the samples as specified for the method.
  2. Fill the sample plates or trays as specified for the method. Waters vials are preferred.
    See Observing vial and plate recommendations.
    Tip: Vial positions V1 through V12, located on the right-hand and left-hand sides of the sample tray, accommodate 4-mL vials. Contact Waters for inserts that allow you to use 2-mL vials in these positions.
  3. Open the sample manager compartment door to access the platter as shown in the figure.
    Figure 1: Sample manager platter
  4. Press the plate/tray selector switch in the lower-left corner of the sample manager as needed to select position 1, 2, or 3.
    Exception: If you press the selector switch while a diagnostic function is running, the sample manager is priming, or the sample needle is accessing the sample tray, making an injection, or being cleaned, the sample manager compartment light blinks continuously and the plate position does not change. The switch operates again after the sample manager completes the task in progress.
  5. Pull out the frame for the selected position by grasping its handle.
  6. Load the plate or tray onto the extended frame. The tray should sit flat.
    Tip: For trays, “A” represents the row and “1” represents the vial position.
    Figure 2: Trays loaded on the sample platter
    Figure 3: Sample plate vial positions
  7. Slide the plate or tray backward until it clicks into place.
    Note: To avoid damaging the sample needle, the sample plates must be positioned correctly, and the sample tray must be fully engaged.
  8. Repeat the previous steps for each plate or tray.
  9. Close the sample compartment door.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

有關最新資訊,請參閱此頁面的英文版。

回到頁首 回到頁首