Short or long 1/4-28 fitting with flangeless ferrule and stainless steel lock ring

Short or long 1/4-28 fitting with flangeless ferrule and stainless steel lock ring
Topic LCI-USG-0067, 最近更新 05/01/2025

Tighten the fitting finger-tight.

Important: Short fittings are intended for use with 1/16-inch OD tubing. Long fittings are intended for use with 1/8-inch OD tubing.

Figure 1: Short 1/4-28 fitting with flangeless ferrule and SST lock ring, first use or reinstallation
Compression screw
Lock ring
Ferrule
End of lock ring with larger ID
Figure 2: Long 1/4-28 fitting with flangeless ferrule and SST lock ring, first use or reinstallation
Compression screw
Lock ring
Ferrule
End of lock ring with larger ID